Accuracy of a novel modified single computed tomography scanning method for assisting dental implant placement: a retrospective observational study

Abstract

Purpose

The aim of this study is to compare dental implant placement accuracy of three surgical guide fabrication methods: single (SCT) and double computed tomography (DCT), and a newly developed modified SCT (MSCT) scan method.

Methods

A total of 183 cases (183 surgical guides, and 485 implants) of static-guide-assisted implant placement surgery using the SCT, DCT, or MSCT methods in a dental clinic were included in the study. Three-dimensional (3D) deviations (mm) at the entry and tip of the implant body between preoperative simulation and actual placement were measured as surrogate endpoints of implant placement accuracy. The following survey details were collected from medical records and CT data: sex, age at implant placement surgery, surgical guide fabrication method, number of remaining teeth, implant length, implant location, alveolar bone quality, and bone surface inclination at implant placement site in preoperative simulation, etc. Risk factors for reducing implant placement accuracy were investigated using generalized estimating equations.

Results

The SCT and DCT methods (odds ratios [ORs] vs. MSCT method: 1.438, 1.178, respectively), posterior location (OR: 1.114), bone surface buccolingual inclination (OR: 0.997), and age at implant placement surgery (OR: 0.995) were significant risk factors for larger 3D deviation at the entry; the SCT (OR: 1.361) and DCT methods (OR: 1.418), posterior location (OR: 1.190), implant length (OR: 1.051), and age at implant placement surgery (OR: 0.995) were significant risk factors for larger 3D deviation at the tip of the implant body.

Conclusions

Implant placement accuracy was better using the MSCT method compared to the SCT and DCT methods.

Background

In oral implant treatment, implant placement position has a significant effect on the safety of oral implant surgery and morphology of the final prosthesis. The determination of the three-dimensional (3D) bone morphology by a preoperative computed tomography (CT) scan and placement of the implant in an accurate position through preplanning with a simulation software program are needed for a successful oral implant surgery [1].

Currently, a static surgical guide, which is manufactured using computer-aided design/computer-aided manufacturing (CAD/CAM) technology based on a preoperative simulation, is widely used to achieve accurate implant placement. Existing static surgical guide fabrication methods are roughly classified into two types: a double computed tomography scan (DCT) method [2] and a single CT scan (SCT) method [3, 4].

The DCT method uses two types of CT images: a 3D CT image obtained with the patient wearing a template fabricated with polymerized acrylic resin and a 3D CT image of the template. In this method, the 3D digital imaging and communications in medicine (DICOM) data from the two images are simultaneously fused with the images of reference markers located in the template vestibule. Using this method, a surgical guide can be fabricated for any situation, from minor tooth loss to complete edentulism. However, the dimensional errors in surgical guides are not negligible in some clinical cases of the DCT method [5] because surgical guides were made from 3D CT images, wherein the threshold setting could affect the shape reproducibility. Block et al. also reported matching errors attributed to artifacts from the markers during CT scanning [6]. In addition, the presence of artifacts can conceal the 3D anatomical structures of soft and hard tissues, which may adversely affect diagnostic accuracy [3, 7].

The SCT method uses one type of CT image. To accurately reflect the preoperative simulation in the surgical guides, highly accurate overlapping 3D CT images (DICOM) and 3D oral plaster model (stereolithography; STL) surface images were obtained with reference to the surface morphology of the remaining teeth. This enabled the fabrication of a surgical guide from the intraoral surface STL data, which were unrelated to the CT-threshold settings. However, the SCT method is not applicable in complete edentulous patients and patients with a small number of remaining teeth owing to challenges in accurate data matching. Furthermore, metal artifacts appear frequently in the region with the remaining teeth on 3D CT images of patients with metal restorations, which reduces the matching point and matching accuracy.

Therefore, to solve all the aforementioned problems, a modified single CT (MSCT) scan method was developed and introduced [8]. The distinctive feature of the MSCT method is the use of a newly developed CT matching template (CTMT) with reference markers made of glass ceramics, which hardly generate artifacts [9]. In this method, three 3D images are superimposed: (1) a CT image (DICOM data) with the patient wearing a CTMT with glass ceramic reference markers, (2) a 3D surface image (STL data) of the patient’s oral plaster model without a CTMT, and (3) the STL data image with a CTMT. The matching accuracy can be improved using this superimposition method as the generated artifacts are extremely small. In addition, the artifacts during CT imaging in the simulation software program can be automatically deleted by a Boolean operation. The MSCT method has the advantages of both the SCT and DCT methods; in specific, it can be applied in all cases, from those involving minor tooth loss to complete edentulous (an advantage of the DCT method) patients, and it can be used to fabricate a surgical guide from high-resolution intraoral surface images (an advantage of the SCT method). However, the accuracy of guided surgery using the MSCT method has not been evaluated in comparison with that of other existing surgical guide fabrication methods.

Therefore, we conducted a retrospective observational study to evaluate the accuracy of computer-guided implant surgery using a surgical guide fabricated by the MSCT method in comparison with using the DCT and SCT methods.

Methods

Cases

The survey included all the cases who underwent static-guide-assisted oral implant placement surgery at Shimizu Dental Clinic (one facility in Japan) from March 1, 2014, to March 1, 2018, using one of the following CT scanning methods: (1) SCT method, (2) DCT method, or (3) MSCT method. One surgeon well-trained for static surgical guides (H.S.) decided the most suitable computer-guided system for each case according to the patient’s oral condition before the surgery. From March 2014 to September 2015, the DCT method was only used. In October 2015, the SCT and MSCT methods were introduced. Thereafter, an appropriate method was chosen from the DCT, SCT, and MSCT methods. The SCT method was only applied to cases with one or more untreated natural teeth (far from artifacts) in each of the 3 portions, anterior teeth, and right and left molars. From December 2016, the SCT and MSCT methods were applied except for the DCT method because the MSCT method gained credibility. The exclusion criteria were as follows: (1) cases in whom the guided surgery system could not automatically match the DICOM data between the pre- and post-operative CT scans and (2) those who did not provide their consent to participate in the study.

The research protocol was approved by the Okayama University Ethics Committee (Ethics Committee No. 14000046, Approval Number: 1806-031). Patients provided written informed consents with permission to use their data for scientific purposes.

Treatment steps

Treatment steps for the SCT method were as follows. A preoperative CT scan (Aquilion Lightning, Canon Medical Systems, Japan) was performed to obtain the DICOM data of maxillofacial region. Definitive impressions were taken using silicone impression materials (ImprintTM4 Penta™ Soft Tray, 3M EPSE, Aquasil Ultra, Dentsply Sirona, USA) to fabricate plaster cast model of intraoral morphology. The obtained plaster cast model was scanned with 3D desktop scanner (CARES® Scanner D7 plus, Straumann, Switzerland), then converted to STL data. The DICOM data of the maxillofacial region and STL data of the plaster cast model were imported to a simulation software program (coDiagnostiX®, Dental Wings Inc., Canada) and superimposed with reference to the surface morphology of the remaining teeth. The implant placement simulation and surgical guide design were performed based on the superimposed data. Surgical guide was designed to fit to the STL data of the intraoral morphology and fabricated using a 3D printer (CARES® P Series P40, Straumann, Switzerland).

In the DCT method, definitive impression was taken using silicone impression materials (ImprintTM4 Penta™ Soft Tray, 3 M EPSE, Aquasil Ultra, Dentsply Sirona, USA) to fabricate the intraoral plaster cast model. A dental technician fabricated a radiographic guide, by burying 6–8 gutta-percha on the plaster cast model. Two types of CT scan, with a radiographic guide and the patient wearing the radiographic guide, were performed (Aquilion Lightning, Canon Medical Systems, Japan), and their DICOM data were obtained. Two types of DICOM data were imported to the simulation software program (Nobel Clinician®, Nobel Biocare, Switzerland) and superimposed with reference to the gutta-percha points. After performing the implant placement simulation and surgical guide design, the surgical guide was fabricated based on the DICOM data of the radiographic guide using the optical shaping method (Nobel Biocare, Switzerland).

The MSCT method was performed according to the protocol previously described by Shimizu et al. [8]. Definitive impression was taken using silicone impression materials (ImprintTM4 Penta™ Soft Tray, 3 M EPSE, Aquasil Ultra, Dentsply Sirona, USA) to fabricate the plaster cast model of the intraoral morphology. The obtained plaster cast model was scanned using a 3D desktop scanner (CARES® Scanner D7 plus Straumann, Switzerland) and converted to STL data. Resin template for the CT scan was printed using a 3D printer (Form 2, Formlabs, USA) based on the obtained STL data of the intraoral morphology. Six glass ceramics markers were added on the occlusal surface of the resin template. Finally, CTMT was completed. CT scan (Aquilion Lightning, CANON MEDICAL SYSTEMS, Japan) of the patient wearing CTMT was performed. Two types of scanning, plaster cast model itself and CTMT attached plaster cast model, were performed (CARES® Scanner D7 plus, Straumann, Switzerland). Then, the obtained STL and DICOM data of the patient wearing the CTMT were imported to the simulation software program (coDiagnostiX®, Dental Wings Inc., Canada) and superimposed with reference to the glass ceramics markers and remaining teeth morphology. After implant placement simulation, the surgical guide design was performed based on the STL data of the intraoral morphology, and the surgical guide was fabricated by a 3D printer (CARES® P Series P40, Straumann, Switzerland).

The surgical guides were fabricated using each methodology, and their internal fittings were confirmed through the pre-formed inspection windows before surgery. During oral implant surgery, all the surgical guides were anchored using fixation pins. The implant surgeries were open-flap or flapless, according to the recommended guided surgery protocol of the manufacturer. Implant bodies were placed without the removal of the surgical guide. A postoperative CT scan was taken to confirm whether the actual implant position was clinically appropriate. Placed implants were selected from following systems: Brånemark System, NobelActive, NobelSpeedy, NobelReplace Straight, NobelReplace/Select Tapered, NobelReplace Conical Connection (Nobel Biocare, Switzerland), and Straumann Bone Level and Bone Level Tapered (Straumann, Switzerland).

Complications

The following events described in the medical records were considered as complications. (1) complications before surgery: (a) failure of the preoperative simulation for designing surgical guides due to artifacts, (2) complications during implant placement surgery: (a) unexpected changes in the implant placement surgery plans (unplanned bone augmentations or changes in implant body diameter or width) during the surgery, (b) fracture of surgical guide templates during the implant surgery (c) unexpected implant body exposure from the alveolar bone surface, (d) perforation into the maxillary sinus or nasal cavity during implant placement surgery, and (e) collision of the implant body with an adjacent tooth.

Observation factors

Alveolar bone quality according to Lekholm and Zarb (type 1/2/3/4) [10] and alveolar bone surface inclination (degree) was evaluated from a preoperative CT image. Bone surface inclination at the implant placement site in preoperative simulation was evaluated separately as buccolingual (bone surface inclination X) and mesiodistal inclination (bone surface inclination Y) (Fig. 1).

Fig. 1
figure 1

Measurement method of alveolar bone surface inclination. a A buccolingual virtual plane (p-1) was set through the implant body axis in the axial slice at the entry of the implant body in the preoperative simulation. b Buccolingual bone surface inclination X (0–90°) was measured between the implant axis and bone surface line on p-1. c A mesiodistal virtual plane (p-2) was set through the implant body axis in the axial slice at the entry of the implant body in the preoperative simulation. d Mesiodistal bone surface inclination Y (0–90°) was measured between the implant axis and bone surface line on p-2

Other information obtained from an electronic medical database for regular implant treatments with reference to the previous reports [11] was as follows: patient’s age at implant placement surgery; sex; surgical guide fabrication method (SCT/DCT/MSCT); number of remaining teeth, coronal teeth (number of remaining teeth with coronal structure and dummy teeth in fixed partial dentures), and teeth with metal restorations (including zirconia restorations); number of placed implants; dentition defect type (Kennedy classification Class I/II/III/IV/complete edentulism); number of fixation pins used during the implant placement surgery; shape of the implant body (straight/tapered); implant width (narrow platform [NP]/regular platform [RP]/wide platform [WP]); implant length; implant location (anterior/posterior, maxillary/mandibular); whether or not immediate implant placement was performed; whether or not bone augmentation was performed; and distance between guide sleeve bottom to bone surface (4.0/5.5/6.0 mm).

Preoperative planning and measurement of 3D deviation between preoperative simulation and actual placement of the implant body (surrogate endpoints)

The planning software program, coDiagnostiX, was used to measure the 3D deviations (mm) at the entry and tip of the implant body between the preoperative simulation and actual placement position by the SCT and MSCT methods. The measurement protocol described by Monaco et al. (2020) was applied to this study [12]. The DICOM data of the pre- and postoperative CT images were assessed using coDiagnostiX, which were adjusted to the same CT threshold and then automatically and accurately superimposed with reference to the characteristic anatomical morphology on CT images using a software function. The pseudo-implant body was accurately placed on the actually placed implant body on the postoperative CT images (Fig. 2a). The 3D positional deviations and distances between the preoperative simulation and actual placement position of the implant bodies were measured automatically using the “Treatment Evaluation” function. The X (mesiodistal axis), Y (buccolingual axis), and Z deviations (depth axis) were automatically measured at the position of entry and tip of the planned and placed implant bodies, and the 3D deviation (3D = \(\sqrt{{\mathrm{X}}^{2}+{\mathrm{Y}}^{2}+{\mathrm{Z}}^{2}}\)) was automatically calculated (Fig. 2b).

Fig. 2
figure 2

Measurement method of the 3D deviation between preoperative simulation and placed implant body position with coDiagnostiX. a A pseudo-implant body was placed based on the actual implant shadow on the CT image after implant placement. b Automatic calculated 3D deviation at the entry and tip between the planned implant body and actual implant body placement position. (Red: Placed implant body; Blue: Planned implant body). 3D, three-dimensional; CT, computed tomography

In the DCT method, the planning software program, Nobel Clinician, was used to measure the 3D deviations at the entry and tip of the implant body between the preoperative simulation and actual placement position. The measurement protocol described by Verhamme et al. (2015) was applied to this study [13]. Superimposition of the pre- and postoperative DICOM data and pseudo-implant body placement were performed using the same method as that used for coDiagnostiX (Fig. 3a). In contrast to the method used with coDiagnostiX, the 3D positional deviations and distances between the actually placed implant body and preoperative simulation were measured manually. The X, Y, and Z deviations at the position of entry and tip between the planned and placed implant bodies were measured with a distance measuring tool, and the 3D deviation was calculated manually using the same formula (Fig. 3b–d).

Fig. 3
figure 3

Measurement method of 3D deviation between preoperative simulation and placed implant body position with Nobel Clinician. a A pseudo-implant body (I-1) was placed based on the actual implant shadow on the CT image after implant placement. b A postoperative pseudo-implant body (I-1) on the cross-sectional slice through the planned implant body (I-2) axis. c Vertices of implant bodies of b were enlarged. 2D deviation of the depth (Z) at the tip between the preoperative and postoperative pseudo-implant body was manually measured using the distance measurement tool. d The axial slice which was contacted to tip of planned implant body (I-2) and rectangular to planned implant (I-2) axis, red line of b. The mesiodistal 2D deviation (X) and buccolingual 2D deviation (Y) at the tip between the preoperative and postoperative pseudo-implant body were manually measured using the distance measurement tool. 3D, three-dimensional; 2D, two-dimensional; CT, computed tomography

Measurements of the 3D deviations between the preoperative simulation and actual placement position of the implant bodies were performed by two examiners (T. M. and Y. K.) independently, while they were not informed about the applied surgical guide fabrication methods. The average of the deviation values measured by the two examiners was adopted as the 3D deviation of each implant.

Statistical analysis

The Kruskal–Wallis and chi-square tests were used to compare the observation factors among the SCT, DCT, and MSCT methods. Since there were no missing data among all observation factors of patients, statistical correction was not needed.

The inter-rater reliability in the 3D deviation measurements of the planned and placed implant bodies between the two examiners was assessed using the Intraclass correlation coefficient (ICC) for each planning software program.

The Steel–Dwass test was used to compare the median 3D deviation between the planned and placed implant bodies among the SCT, DCT, and MSCT methods. Kruskal–Wallis test, Mann–Whitney U test, or Spearman’s rank correlation coefficient was used to analyze the relationships between the predictor variables and the 3D deviations between the planned and placed implant positions at the entry and tip of the implant body between each surgical guide fabrication method.

To evaluate the accuracy of implant placement position using each surgical guide fabrication method, generalized estimating equations (GEEs) were used to identify the risk factors for large 3D deviations at the entry and tip of the implant body between the preoperative simulation and actually placed position. GEEs were implemented using the forced entry method.

Statistical analyses were performed using SPSS for Windows (version 25 for SPSS, IBM, Japan). Steel–Dwass test was performed using (JMP version 11, SAS, Japan). The level of significance was set at p < 0.05.

Results

Cases and complications

A total of 183 cases (158 patients) met the selection criteria and underwent implant surgeries (average age at implant placement surgery: 62.3 ± 11.5 years, male/female: 54/104, 183 surgical guides, and 485 implants).

Of the 183 surgical guides, five DCT surgical guides, involving nine implant bodies, induced complications. The complications were as follows: (1) inability to fabricate the surgical guide due to artifacts (one surgical guide, one implant), (2) collision with adjacent teeth due to displacement of the implant drilling position during the guided surgery (three surgical guides, four implants), and (3) application of bone augmentation for unexpected bone defects due to drilling errors (one surgical guide, four implants). Of the three cases (five surgical guides) with complications, cases (1) and (3) were excluded from the analyses as the pre- and postoperative CT data could not be superimposed.

The number of analyzed cases was 181 (156 patients, average age at implant placement surgery: 61.7 ± 11.8 years, male/female: 54/102, 181 surgical guides, and 480 implants). Table 1 shows the basic data for each surgical guide fabrication method. Significant differences were observed between the surgical guide fabrication methods, except for in terms of sex, age at implant placement surgery, number of teeth with metal restorations, implant location (anterior/posterior), and presence or absence bone augmentation (Table 1).

Table 1 Basic data of each surgical guide fabrication method

Comparison of 3D deviation between planned and placed implant position (univariate analysis)

The 3D deviations between the planned and placed implant positions at the entry and tip of the implant body between each surgical guide fabrication method were compared using the Steel–Dwass test. As a result, the median 3D deviations of the SCT method at the entry was 0.788 mm (first quartile: 0.623, third quartile: 1.008) and that at the tip was 1.084 mm (first quartile: 0.742, third quartile: 1.269). The median 3D deviation of the DCT method at the entry was 0.646 mm (first quartile: 0.428, third quartile: 0.891) and that at the tip was 1.100 mm (first quartile: 0.727, third quartile: 1.433). The median 3D deviations of the MSCT method at the entry was 0.522 mm (first quartile: 0.338, third quartile: 0.738) and that at the tip was 0.674 mm (first quartile: 0.451, third quartile: 0.991). The results showed that the median 3D deviations of the MSCT method (entry: 0.522, tip: 0.674) was significantly smaller than that of the SCT (entry: 0.788, tip: 1.084) and DCT (entry: 0.646, tip: 1.100) methods at both the entry and tip of the implant body (Table 2, p < 0.01).

Table 2 Comparison of 3D deviation by each surgical guide fabrication method

The ICC for inter-examiner reliability was 0.949 for coDiagnostiX and 0.951 for Nobel Clinician. According to Landis and Koch, both reliability levels can be estimated as “almost perfect” [14].

Examination of factors affecting implant accuracy (multivariate analysis with GEEs)

The following predictor variables were submitted initially as the factors related to each surgical guide: surgical guide fabrication method (SCT/DCT/MSCT), sex (male/female), age at implant placement surgery, number of coronal teeth, number of teeth with metal restorations, number of placed implants, dentition defect type (Kennedy Classification Class I/II/III/IV/complete edentulism), and number of fixation pins used during surgery; further, the following were submitted as the factors related to each implant body: implant body shape (straight/tapered), width (NP/RP/WP), length, and location (anterior/posterior, maxillary/mandibular); immediate implant placement (with/without); bone augmentation (with/without); alveolar bone quality; distance between the guide sleeve bottom to bone surface; and bone surface inclination (X, Y).

The relationships between the predictor variables and the 3D deviations between the planned and placed implant positions at the entry and tip of the implant body among each surgical guide fabrication methods are shown in Table 3.

Table 3 The relationships between predictor variables and the 3D deviations between the planned and placed implant positions at the entry and tip of the implant body among each surgical guide fabrication methods

Table 4 shows the results of the GEE analysis for the risk factors for large 3D deviations at the entry of the implant body between the preoperative simulation and actual placement position. The SCT (p < 0.001) and DCT (p = 0.006) methods were compared to the MSCT method; posterior implant position (p = 0.034), smaller bone surface inclination X (p = 0.006), and younger age at implant placement surgery (p = 0.036) were significant risk factors for large 3D deviations.

Table 4 Risk factors for large 3D deviation at the entry of the implant body between preoperative simulation and actual placement analyzed by GEEs

Table 5 shows the results of the GEE analysis for the risk factors for 3D deviations of the tip of the implant body. The SCT method (p = 0.003) and DCT method (p < 0.001) were compared to MSCT method, posterior implant location (p = 0.001), longer length of the implant body (p = 0.006), and younger age at implant placement surgery (p = 0.043) were significant risk factors for large 3D deviations.

Table 5 Risk factors for large 3D deviation at the tip of the implant body between preoperative simulation and actual placement analyzed by GEEs

Discussion

This retrospective observational study aimed to evaluate the differences in the accuracy of the implant placement position between the newly developed MSCT method and existing surgical guide fabrication methods. Most of the previous studies, which evaluated the effect of the surgical guide system on implant placement accuracy, focused on comparisons between accuracy with and without a surgical guide [15]. None of the studies compared the placement accuracies between different surgical guide fabrication methods. Therefore, this is the first study to include the types of static surgical guide fabrication methods as explanatory variables and to identify the risk factors that might reduce implant placement accuracy. The details of each issue investigated in this study are discussed below.

Comparison of 3D deviation between planned and placed implant positions (univariate analysis)

Inter-examiner reliability in evaluating the 3D deviation between the preoperative simulation and actual implant placement positions was assessed. The ICC of both methods, using coDiagnostiX and Nobel Clinician, was more than 0.9, which indicates that both measurement methods are sufficiently reliable.

Regarding the DCT method, Van Assche et al. (2012) reported that the mean 3D deviation at the entry of the implant body was 0.99 mm and that at the tip of the implant body was 1.24 mm [16]. According to a meta-analysis of 3D deviations at the implant body placement position on using the DCT and SCT methods, the mean 3D deviations were 1.2 and 1.4 mm at the entry and tip, respectively [17]. In this study, the mean 3D deviations of the SCT method was 0.82 mm at the entry and 1.00 mm at the tip and that of the DCT method was 0.69 mm at the entry and 1.13 mm at the tip. Taking these into consideration, implant placement surgeries with the SCT and DCT methods in this study produced equivalent or higher accuracies than those previously reported. Nevertheless, the MSCT method showed significantly smaller 3D deviations than the SCT and DCT methods at both the entry and tip of the implant body. These results indicate the effectiveness of the MSCT method. However, since this study was an observational study, it must be noted that the baseline demographic data of patients in each study group based on the applied surgical guide were different. Since there were significant differences in the distribution of the baseline data between the three groups, the difference in 3D deviations between them might not have originated from the accuracy of the surgical guide fabrication method. Future studies that address this drawback are necessary.

Factors related to implant placement accuracy (multivariate analysis with GEEs)

The predictors employed in this study were factors related to both the surgical guides and implant bodies. Thus, we utilized GEEs in this study, which can be used to analyze the effects of both categories of related factors separately, instead of a multiple regression analysis. We considered that employing this statistical analysis method would be substantially advantageous because the mean number of implants placed per surgical guide was 2.9 ± 1.9, implying that one surgical guide was prepared for the placement of multiple implant bodies in this study.

According to the results of GEEs, the application of the MSCT method can significantly reduce the 3D deviations between pre- and postoperative implant positions considering the surgical guide fabrication methods. A possible reason for the reduction in the 3D deviation was the difference in the imaging matching procedure employed in the MSCT method. As mentioned above, we utilized newly developed reference markers made of glass ceramics, which reduce the production of artifacts in the matching process between the 3D CT image and 3D intraoral surface shape image in the MSCT method. By contrast, both the images were superimposed based on the 3D images of the remaining teeth in the SCT method. However, with this procedure, matching errors may occur owing to the artifacts generated from metal restorations on these teeth [18]. In cases involving the DCT method, artifacts generated from the markers for image matching may induce matching errors. In addition, the fact that surgical guides can be produced by the morphology built from 3D CT image to be influenced by CT-threshold settings can be considered to be related to implant placement accuracy in these cases.

Derksen et al. (2019) reported that 3D deviations at both the entry and tip of the implant body significantly increased in cases with 5–6 unrestored remaining teeth compared to in those with 7 or more unrestored remaining teeth when using the SCT method [11]. In addition, Nabha et al. (2014) demonstrated that the greater the number of restored remaining teeth, the greater the number of generated metal artifacts; this causes matching errors between the 3D CT image and 3D surface image of the remaining teeth (STL) [19]. However, the numbers of remaining teeth with coronal structures and metal restorations were not identified as the significant risk factors for large 3D deviations in this study. A possible reason for this phenomenon was that this study included patients who underwent procedures using the MSCT and DCT methods, which are not affected by the artifacts caused by metal restorations during the matching process.

Implant placement in the posterior location was a significant risk factor for increased 3D deviation at the entry and tip of the implant body. When placing implants in the posterior region, confirming whether the hand drill is positioned correctly to fit the guide sleeve against the anterior region could be more challenging. In addition, drill entry angles are more susceptible to restriction due to the shorter distance between the alveolar crest at the placement site and opposing teeth. Thus, 3D deviations might increase when implants are placed in the posterior region. Furthermore, placement of a longer implant body was a significant risk factor for increased 3D deviation at the tip of the implant body. When drilling is angled, the 3D deviations at the tip between the preoperative simulation and postoperative implant placement sites could increase with longer implant placement. This finding is consistent with that in a study by D’haese et al. (2012), which utilized the NobelGuide system [20].

Larger alveolar bone surface inclination at the placement site was a significant risk factor for an increase in 3D deviations at the entry of the implant body. A previous report suggested a certain degree of variable range between a guide sleeve and drill, which can cause an error between preoperatively planned and actual implant placement sites [21]. Moreover, a mechanical engineering report by Sakuma et al. (1983) demonstrated that a drill hole was formed at a deviation from the target position during drilling on an inclined surface [22]. Once the drill hole was deviated, they concluded that the amount of cumulative displacement could not be corrected. In addition, Ohnishi et al. (2003) reported that a disproportion of the radial force at the starting point of drill penetration occurred on inclined surfaces, thereby decreasing the drilling accuracy [23]. Hence, the steep inclination of the alveolar bone surface might cause drilling and implant placement errors owing to the sliding of the drill apex and implant body from the target position, causing subsequent changes in direction. Further studies are required to confirm these findings. Further, age at implant placement surgery was identified as a risk factor for an increase in 3D deviations at the tip and the entry of the implant body. This had no multicollinearity with other observational factors; thus, it was completely independent. Previous reports showed that cancellous bone tissue structure became sparse with aging [24] and cancellous bone became low density and more porous [25]. Thus, younger people’s hard bone tissue structure and their high bone density might affect to these results.

Suspected observation factors, which could affect implant placement accuracy, were not identified as significant risk factors, e.g., coronal teeth, dentition defect type, number of fixation pins, and distance between sleeve bottom to bone surface according to the GEE analyses. However, it should not be interpreted that these factors never affect implant placement accuracy. According to the basic data, significant correlations or differences were observed between the surgical guide fabrication methods and these suspected observation factors. Considering these relationships, the GEE analyses might have been affected by multicollinearity. Thus, the surgical guide fabrication method, which was the most affected factor among mutually related observation factors, was only identified as a significant risk factor for implant placement accuracy.

Complications

In this study, severe complications, such as a collision of the placed implant body with an adjacent tooth, unexpected changes in the implant surgery plan, and unexpected bone augmentation procedures due to drilling errors during surgeries, were seen only in the cases where the DCT method was applied. In this study, the median values of the 3D deviation in the DCT group were almost identical to those in the SCT group. However, the range from minimum to maximum 3D deviations in the DCT group was wider than that in the SCT group. In addition, the DCT method was selectively applied to cases with large edentulous spaces in this study. In these cases of large edentulous areas, when the surgical guide is placed in the oral cavity, it is generally difficult to confirm whether it is precisely restored to the planned position. Therefore, the occurrence of severe complications was observed in the DCT group.

Limitations

In this study, two examiners measured the 3D deviations between the preoperatively planned and actual implant positions in each case, while not being informed about the surgical guide fabrication method. However, the examiners could predict the implant system and type of surgical guide in the cases where the DCT method was applied because Nobel Clinician, the planning software program, was only used for the measurement of DCT method. Two type of software programs, coDiagnostiX and Nobel Clinician, were used to measure the outcomes of this study. Both the methodologies were utilized in previously reported outcome [12, 13, 26, 27] measurements but the degree of coincidence between the two has not been evaluated. Even though this research was a retrospective study, there were no missing data because the clinic that conducted this study had regularly constructed the clinical database for implant treatment. This enabled compensation of the shortcomings of retrospective studies. However, since every surgery was performed by one operator (H.S.), there was a limitation in generalizing the findings of this study.

Conclusions

This retrospective observational study demonstrated that the MSCT method significantly improves the implant placement accuracy at both the entry and tip of the implant body, compared with the DCT and SCT methods. Severe complications were observed only in the cases where the DCT method was applied.

Availability of data and materials

The datasets generated and analyzed during the current study are available from the corresponding author on reasonable request.

Abbreviations

3D:

Three-dimensional

SCT:

Single computed tomography

DCT:

Double computed tomography

MSCT:

Modified single computed tomography

DICOM:

Digital imaging and communications in medicine

STL:

Stereolithography

CTMT:

Computed tomography matching template

NP:

Narrow platform

RP:

Regular platform

WP:

Wide platform

ICC:

Intraclass correlation coefficient

GEE:

Generalized estimating equation

OR:

Odds ratio

References

  1. Sarment DP, Sukovic P, Clinthorne N. Accuracy of implant placement with a stereolithographic surgical guide. Int J Oral Maxillofac Implants. 2003;18(4):571–7.

    PubMed  Google Scholar 

  2. Sun Y, Luebbers HT, Agbaje JO, Schepers S, Politis C, Van Slycke S, et al. Accuracy of dental implant placement using CBCT-derived mucosa-supported stereolithographic template. Clin Implant Dent Relat Res. 2015;17:862–70. https://doi.org/10.1111/cid.12189.

    Article  PubMed  Google Scholar 

  3. Flügge TV, Nelson K, Schmelzeisen R, Metzger MC. Three-dimensional plotting and printing of an implant drilling guide: simplifying guided implant surgery. J Oral Maxillofac Surg. 2013;71(8):1340–6. https://doi.org/10.1016/j.joms.2013.04.010.

    Article  PubMed  Google Scholar 

  4. Kernen F, Benic GI, Payer M, Schär A, Müller-Gerbl M, Filippi A, et al. Accuracy of three-dimensional printed templates for guided implant placement based on matching a surface scan with CBCT. Clin Implant Dent Relat Res. 2016;18(4):762–8. https://doi.org/10.1111/cid.12348.

    Article  PubMed  Google Scholar 

  5. Verhamme LM, Meijer GJ, Boumans T, Schutyser F, Bergé SJ, Maal TJ. A clinically relevant validation method for implant placement after virtual planning. Clin Oral Implants Res. 2013;24(11):1265–72. https://doi.org/10.1111/j.1600-0501.2012.02565.x.

    Article  PubMed  Google Scholar 

  6. Block MS, Chandler C. Computed tomography-guided surgery: complications associated with scanning, processing, surgery, and prosthetics. J Oral Maxillofac Surg. 2009;67(11 Suppl):13–22. https://doi.org/10.1016/j.joms.2009.04.082.

    Article  PubMed  Google Scholar 

  7. Park HH, Shin JY, Lee J, Jin GH, Kim HS, Lyu KY, et al. A study on the artifacts generated by dental materials in PET/CT image. Annu Int Conf IEEE Eng Med Biol Soc. 2013. https://doi.org/10.1109/EMBC.2013.6610039.

    Article  PubMed  Google Scholar 

  8. Shimizu H, Arakawa H, Mino T, Kurosaki Y, Tokumoto K, Kuboki T. Newly developed data-matching methodology for oral implant surgery allowing the automatic deletion of metal artifacts in 3D-CT images using new reference markers: a case report. J Prosthodont Res. 2021;65(1):125–9. https://doi.org/10.2186/jpr.JPOR_2019_429.

    Article  PubMed  Google Scholar 

  9. Rodrigues SP, Paiva JM, De Francesco S, Amaral MI, Oliveira FJ, Silva RF. Artifact level produced by different femoral head prostheses in CT imaging: diamond coated silicon nitride as total hip replacement material. J Mater Sci Mater Med. 2013;24(1):231–9. https://doi.org/10.1007/s10856-012-4778-0.

    Article  PubMed  Google Scholar 

  10. Brånemark PI, Zarb GA, Albrektsson T. Tissue-integrated prosthesis. Chicago: Quintessence Publishing; 1985.

    Google Scholar 

  11. Derksen W, Wismeijer D, Flügge T, Hassan B, Tahmaseb A. The accuracy of computer-guided implant surgery with tooth-supported, digitally designed drill guides based on CBCT and intraoral scanning a prospective cohort study. Clin Oral Implants Res. 2019;30(10):1005–15. https://doi.org/10.1111/clr.13514.

    Article  PubMed  Google Scholar 

  12. Monaco C, Arena A, Corsaletti L, Santomauro V, Venezia P, Cavalcanti R, et al. 2D/3D accuracies of implant position after guided surgery using different surgical protocols: a retrospective study. J Prosthodont Res. 2020;64(4):424–30. https://doi.org/10.1016/j.jpor.2019.11.007.

    Article  PubMed  Google Scholar 

  13. Verhamme LM, Meijer GJ, Berg SJ, Soehardi RA, Xi T, de Haan AF, et al. An accuracy study of computer-planned implant placement in the augmented maxilla using mucosa-supported surgical templates. Clin Implant Dent Relat Res. 2015;17(6):1154–63. https://doi.org/10.1111/cid.12230.

    Article  PubMed  Google Scholar 

  14. Landis JR, Koch GG. The measurement of observer agreement for categorical data. Biometrics. 1977;33(1):159–74.

    Article  PubMed  Google Scholar 

  15. Chen S, Ou Q, Lin X, Wang Y. Comparison between a computer-aided surgical template and the free-hand method: a systematic review and meta-analysis. Implant Dent. 2019;28:578–89. https://doi.org/10.1097/ID.0000000000000915.

    Article  PubMed  Google Scholar 

  16. Van Assche N, Vercruyssen M, Coucke W, Teughels W, Jacobs R, Quirynen M. Accuracy of computer-aided implant placement. Clin Oral Implants Res. 2012;23(Suppl 6):112–23. https://doi.org/10.1111/j.1600-0501.2012.02552.x.

    Article  PubMed  Google Scholar 

  17. Tahmaseb A, Wu V, Wismeijer D, Coucke W, Evans C. The accuracy of static computer-aided implant surgery: a systematic review and meta-analysis. Clin Oral Implants Res. 2018;29(Suppl 16):416–35. https://doi.org/10.1111/clr.13346.

    Article  PubMed  Google Scholar 

  18. Flügge T, Derksen W, Te Poel J, Hassan B, Nelson K, Wismeijer D. Registration of cone beam computed tomography data and intraoral surface scans—a prerequisite for guided implant surgery with CAD/CAM drilling guides. Clin Oral Implants Res. 2017;28(9):1113–8. https://doi.org/10.1111/clr.12925.

    Article  PubMed  Google Scholar 

  19. Nabha W, Hong YM, Cho JH, Hwang HS. Assessment of metal artifacts in three-dimensional dental surface models derived by cone-beam computed tomography. Korean J Orthod. 2014;44(5):229–35. https://doi.org/10.4041/kjod.2014.44.5.229.

    Article  PubMed  PubMed Central  Google Scholar 

  20. D’haese J, Van De Velde T, Elaut L, De Bruyn H. A prospective study on the accuracy of mucosally supported stereolithographic surgical guides in fully edentulous maxillae. Clin Implant Dent Relat Res. 2012;14(2):293–303. https://doi.org/10.1111/j.1708-8208.2009.00255.x.

    Article  PubMed  Google Scholar 

  21. Koop R, Vercruyssen M, Vermeulen K, Quirynen M. Tolerance within the sleeve inserts of different surgical guides for guided implant surgery. Clin Oral Implants Res. 2013;24(6):630–4. https://doi.org/10.1111/j.1600-0501.2012.02436.x.

    Article  PubMed  Google Scholar 

  22. Sakuma K, Taguchi K, Katsuki A. Axial deviation of hole in deep drilling (1st Report)—influence of misalignments in drilling system. J Jpn Soc Precis Eng. 1983;49(10):1379–84. https://doi.org/10.2493/jjspe1933.49.1379.

    Article  Google Scholar 

  23. Ohnishi O, Onikura H. Effects of ultrasonic vibration on microdrilling into inclined surface. J Jpn Soc for Precis Eng. 2003;69:1337–41. https://doi.org/10.2493/jjspe.69.1337.

    Article  Google Scholar 

  24. Giordano V, Franco JS, Koch HA, Labronici PJ, Pires RE, Amaral NP. Age-related changes in bone architecture. Rev Col Bras Cir. 2016;43(4):276–85. https://doi.org/10.1590/0100-69912016004010.

    Article  PubMed  Google Scholar 

  25. Skrzat J, Brzegowy P, Walocha J, Wojciechowski W. Age dependent changes of the diploe in the human skull. Folia Morphol (Warsz). 2004;63(1):67–70.

    PubMed  Google Scholar 

  26. Feng Y, Su Z, Mo A, Yang X. Comparison of the accuracy of immediate implant placement using static and dynamic computer-assisted implant system in the esthetic zone of the maxilla: a prospective study. Int J Implant Dent. 2022;8(1):65. https://doi.org/10.1186/s40729-022-00464-w.

    Article  PubMed  PubMed Central  Google Scholar 

  27. Stünkel R, Zeller AN, Bohne T, Böhrnsen F, Wedi E, Raschke D, Kauffmann P. Accuracy of intraoral real-time navigation versus static, CAD/CAM-manufactured pilot drilling guides in dental implant surgery: an in vitro study. Int J Implant Dent. 2022;8(1):41. https://doi.org/10.1186/s40729-022-00430.

    Article  PubMed  PubMed Central  Google Scholar 

Download references

Acknowledgements

H.S. holds patents (P6238330) on the newly developed reference marker and matching workflow.

Funding

The authors declare that they have not received any funding.

Author information

Authors and Affiliations

Authors

Contributions

HS collected the data and performed the treatments; TM and HA conceived the study design and methodology; TM and YK measured the outcomes; KT analyzed the data; TM and AK interpreted the results; HS and KT wrote the draft; TM and KM wrote the manuscript; TK supervised the study and wrote the manuscript. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Takuya Mino.

Ethics declarations

Ethics approval and consent to participate

The research protocol was approved by the Okayama University Ethics Committee (Ethics Committee No. 14000046, Approval Number: 1806-031).

Consent for publication

Informed consent from all patients of the present study was obtained, respectively.

Competing interests

The authors declare that no conflicts of interest.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Shimizu, H., Mino, T., Kurosaki, Y. et al. Accuracy of a novel modified single computed tomography scanning method for assisting dental implant placement: a retrospective observational study. Int J Implant Dent 9, 42 (2023). https://doi.org/10.1186/s40729-023-00509-8

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s40729-023-00509-8

Keywords