Implant primary stability depending on protocol and insertion mode — an ex vivo study

Abstract

Background

Dental implant primary stability is thought to be a fundamental prerequisite for the long-term survival and success. The aim of this study was to analyze the influence of protocol and insertion mode on dental implant stability ex vivo. One hundred and twenty implants were inserted either manually or machine-driven into porcine mandibles by a standard or over-dimensioned protocol. Dental implant stability was measured via resonance frequency analysis (RFA), insertion torque (IT), and torque out (TO).

Results

Statistically significant higher IT and TO values were seen after standard protocol insertion (p < 0.05), whereas manual and machine-driven insertion mode showed equivalent values.

Conclusions

The over-dimensioned protocol exceeded the primary stability values recommended for immediate implant insertion; therefore, it could be recommended as well.

Background

A reliable option for replacing teeth is the insertion of osseointegrated implants. Dental implant primary stability (DIS) has also been reported to be a fundamental prerequisite for long-term success of dental implants [1, 2], even though osseointegration has also been achieved without a certain amount of primary stability [3, 4]. Primary stability has been defined as the ability to withstand axial, lateral, and rotational loading [5] and depends on the implants’ anchorage within the bone [6]. The creation of different implant geometries and drilling protocols seems to have improved the achievement of high primary stability within the bone. The interaction between the implants’ geometric properties, combined with the surgical drilling technique indicated for the detected bone density can contribute to obtaining low amounts of compressive stress and micromotions on the surrounding bone tissue during placement [6,7,8].

For evaluation of in vivo primary stability, insertion torque (IT) and resonance frequency analysis (RFA) are well-established methods [2, 9]. IT is a mechanical parameter influenced by surgical procedure [10, 11], implant design [8, 10, 11], and bone quality [2, 6, 8, 10,11,12]. It has been defined by the Foundation for Oral Rehabilitation as the cutting resistance of the bone during implant insertion, the friction, and has been considered an indirect value of implant primary stability [13]. Values above 32 Ncm indicate that the implant is firmly embedded in the bone and mechanically stable [2, 4, 14]. RFA assumes that the frequency is directly related to the stiffness of the bone-implant interface and the surrounding bone [5, 14, 15]. High values show a stable implant and allow verification of osseointegration and secondary stability over time [11, 16]. RFA is measured by Ostell®-devices (Osstell ISQ, Osstell, Göteborg, Sweden). A sensor (SmartPeg, Osstell, Göteborg, Sweden) is mounted on the implant and is vibrated by moving it with magnetic pulses. With increasing stiffness of the bone-implant interface, the vibration frequency of the sensor increases. While RFA is expressed in hertz, implant stability quotient (ISQ) is the scale used to quantify RFA values (range 1–100) [14, 17]. Even though RFA has been reported to be a reliable, reproducible, and objective method to measure the stiffness of bone-implant-complex [11, 18], it has also been reported that RFA data from immediately placed implants could be misleading in increasing in terms of predicting primary stability [19,20,21]. Therefore, this value needs to be supported by another means of quantifying anchorage, especially for immediate loading protocols [18]. The reverse torque test or torque out (TO) could be used in vitro/ex vivo, as it gives information on the ability of the abutment or the whole implant to withstand a given torque value [10, 14, 22,23,24,25]. As this test means to unscrew the implant, this method is inappropriate for clinical use.

Patient-dependent factors affecting implant stability include bone quality and quantity. Greater stability was achieved in more dense bone [1, 2, 4, 6, 8, 10, 12, 14, 15, 18]; here, failure rates accumulated in the upper jaw [2, 12, 15]. Age, gender, smoking status, and periodontal status also affect the DIS [15]. Implant-related factors affecting DIS vary from implant macro-geometry to surface characteristics [1, 4]. The surgical technique also influences DIS [6, 12, 14], and higher insertion torques as well as under-dimensioned drilling protocols are thought to increase the percentage of initial bone-implant contact by a better fit of the implant into the bone [4, 18, 26] which in turn reduces the amount of micromotion after implant insertion [7, 8, 10, 27].

Dental implant failure is related to numerous factors such as the quality and quantity of the bone-implant interface [4, 8, 14, 18], the type of loading [1, 4, 7, 8, 12, 14, 18, 26], implant geometry [1, 4,5,6, 8, 12, 15, 18, 26, 28] and surface characteristics [4, 14, 28], restoration prosthesis type, and the surgeons’ experience [8]. Even though previous studies showed higher stability values for implants inserted in under-dimensioned cavities [7, 18], in the last few years research has been directed to over-dimensioned protocols (ODP). Analyzing those, a decrease in primary and an increase in secondary stability with a shorter healing period for implants became apparent [7, 29]. Kim et al. compared the effect of oversized drilling sockets regarding bone-to-implant contact and bone density after 4 and 8 weeks in an in vivo dog model. They used a final drill of 4.00 mm for implants with a diameter of 4 mm in the oversized group and a final drill of 2.85 mm in the control group. The analysis could show that while the initial bone-to-implant contact at 4 weeks was lower in the oversized group, it increased within the following 4 weeks up to values of 77.38% which was higher than the values in the control group (69.52%). After 8 weeks, the mean bone density was shown to be comparable between both groups. No difference could be observed regarding the healing period [30]. Within a study analyzing the differences between under-, over-dimensioned, and intermediate-sized drilling sockets in dogs after a healing period of 2 weeks, Campos et al. could show that although the undersized drilling condition led to the highest insertion torque, new bone formation was most active in the intermediate-sized group. They concluded that high insertion torque values do not necessarily result in the most favorable biologic response [31]. The purpose of this study was to examine the effect of either standard (SP) or ODP, using three different means of quantifying dental implant primary stability. To our best knowledge, only one study evaluated the effect of manual and machine-driven insertion mode so that further analysis was done to quantify the hypothesized differences [26].

The research hypotheses were that a manual insertion mode is more advantageous and that implants inserted by ODP were expected to show a decrease in dental implant primary stability, as described previously [32].

Methods

Bone specimens

Twenty mandibles from fresh porcine cadavers were obtained from a local slaughterhouse. The animals did not show any macroscopic signs of any pathologic bone conditions. After removal of the surrounding soft tissue, the surfaces of the bone samples were thoroughly cleaned. Each sample was checked macroscopically for irregularities and a minimum thickness of 20 mm at the place of the intended implant placement was verified. All obtained samples were immediately used.

Surgery

In the following, a preparation wider than recommended by the company (also if narrower than the implants’ diameter) will be termed as over-dimensioned protocol (ODP).

Osteotomies were done following the manufacturer’s protocol, using sequences of varied diameter drills. One hundred and twenty implants (HiTec Tapered Self Thread implant, Hi-Tec Implants, Herzlia, Israel; 8 mm length and 3.3/3.75 diameter) were inserted either manually or machine-driven into porcine mandibles using SP or ODP. A manual insertion mode was used for implants inserted via SP (n = 45) or ODP (n = 30). A machine-driven mode was used for additional implants inserted via SP (n = 45). Drilling was conducted at 850 rpm and the implant was inserted at a speed of 15 rpm using a surgical motor for dental implants (Elcomed, type SA-310, W&H Dentalwerk Bürmoos GmbH, Bürmoos, Austria). Maximum insertion torque was set at 45 Ncm when implants were machine-driven, whereas manually inserted implants were drilled in by a hand ratchet at max. 45 Ncm.

Preparation protocol for standard osteotomy (SP)

Osteotomies were conducted in accordance with the manufacturer’s protocol [33]. After exposure, the bone was penetrated to a depth equivalent to the implant length using the 2-mm internal irrigation drill. Using the 2.5-mm and 2.8-mm drill, the preparation for 3.3 mm implants was completed. To achieve the space for 3.75 mm implants, the 3.2-mm drill was also used (Fig. 1).

Fig. 1
figure 1

Standard protocol. This figure shows the implant types and drilling protocol used within this study. Standard protocol was conducted by a final drill of 2.80 mm for 3.3 mm implants, 3.20 mm for 3.75 mm implants, and 3.65 mm for 4.2 mm implants. Permissions for reproducing the figures were received from HI-TEC IMPLANTS LTD. Source: Product Catalogue 12th Edition [40]

Preparation protocol for oversized osteotomies (ODP)

This protocol repeated the steps of the standard protocol but then added a larger final drill. For the 3.3-mm implants, the final drill size was 3.2 mm; for the 3.75-mm implants, the final drill size was 3.65 mm (Fig. 2).

Fig. 2
figure 2

Over-dimensioned protocol. The over-dimensioned protocol was conducted by a final drill of 1 mm narrower than the implant diameter. The final drill for implants of 3.3. mm was 3.2 mm and of implants measuring 3.75 mm, it was 3.65 mm. Within this study, an over-dimensioned protocol was defined as a final drill larger than recommended by the company, which is in this case 4 or 4.5 mm wider than used in the standard protocol. Permissions for reproducing the figures were received from HI-TEC IMPLANTS LTD. Source: Product Catalogue 12th Edition [34]

RFA

To analyze the data, an Osstell® SmartPeg threaded transducer (implant diameter 3.3 and 3.75 mm: SmartPeg Type 32 with a platform of 3.5 mm, Art-Nr. 100440; implant diameter 4.2 mm: SmartPeg Type 27 with a platform of 4.5 mm, Art-Nr. 10043; Osstell, Göteborg, Sweden) was screwed onto the implant fixture with the aid of a transporter. RFA was measured by Osstell® electronic analyzer (Osstell ISQ), positioned at a distance of 3 mm from the abutment. The recorded frequency was automatically converted into implant stability quotient (ISQ) values with a range of 0 to 100 (minimum to maximum stability). The mean ISQ values were calculated.

Torque in and torque out

Insertion torque and torque out were measured by using a customized handheld torque screwdriver with a digital output device (Mecmesin, Schwenningen, Germany) as described by Bolm et al. [35]. Torque values were recorded in Newton centimeter.

Statistics

To analyze the differences between the measured value normality and homogeneity of variance tests (Levene statistic) were performed at first in order to check the conditions for the subsequent analysis. p values were obtained with either independent-samples t test or one-way ANOVA. In case of inhomogeneity of variance, a Welch-ANOVA was used instead. To obtain p values, post hoc multiple comparison Tukey test (equal variances assumed) or Games-Howell (equal variances not assumed) was used. The effect size was calculated as described in “Statistical power analysis for the behavioral sciences” by Jacob Cohen in 1988 [36]. Values are displayed as Cohen’s d and effect size (r). The statistical analyses were performed using SPSS version 24 for Windows (IBM, Armonk, New York). A p value 0.05 was termed significant. Values are displayed as mean plus standard deviation.

Results

Drilling protocol: standard versus over-dimensioned

No statistically significant difference in RFA could be measured (Cohen’s d = − 0.022, effect size r = 0.011, p = 0.260), whereas IT values were significantly higher in implants inserted via SP (90.56 ± 31.27 Ncm) in comparison with the ODP (63.74 ± 48.61 Ncm, p = 0.002; Cohen’s d = 0.656, effect size r = 0.312). The analyzed TO values showed similar results with higher values in the SP (93.59 ± 32.3 Ncm) compared with the ODP (58.35 ± 40.43 Ncm, p = 0.043; Cohen’s d = 0.963, effect size r = 0.434) (see Table 1 and Fig. 3).

Table 1 Comparison of standard and over-dimensioned protocol
Fig. 3
figure 3

Comparison of standard and over-dimensioned protocol. The figure displayed shows the comparison between standard and over-dimensioned protocol. a Displays the measurements obtained by RFA. The unit is ISQ with a range of 0 to 100 (minimum to maximum stability). b Displays the results obtained by the torque in and c by the torque out test. Although, there was no statistically significant difference in ISQ between the groups, torque in and torque out tests showed significantly lower values in the over-dimensioned group compared to implant inserted via standard protocol. ISQ, implant stability quotient; IT, insertion torque; TO, torque out

Insertion mode: manual versus machine-driven, standard protocol

No statistically significant difference of RFA (Cohen’s d = –0.309, effect size r = –0.153, p = 0.185), IT (Cohen’s d = 0.21, effect size r = 0.104, p = 0.937), and TO (Cohen’s d = 0.109, effect size r = 0.054, p = 0.490) could be shown between manual and machine-driven insertion mode (see Table 2 and Fig. 4).

Table 2 Comparison of manual and machine-driven insertion mode
Fig. 4
figure 4

Comparison of manual and machine-driven insertion mode. This figure displays the comparison between manual and machine-driven insertion mode. a Displays the results obtained by the RFA test, indicated as ISQ (0–100). b Displays the measurements obtained by the torque in and c by the torque out test. There was no statistically significant difference between the groups. ISQ, implant stability quotient; IT, insertion torque; TO, torque out

Implant geometry: standard protocol

No statistically significant difference regarding ISQ (p = 0.353), IT (p = 0.099), or TO (p = 0.337) could be measured between implants of different diameter. Neither was there a statistically significant difference between implants of different lengths (ISQ: p = 0.164, IT: p = 0.303, TO: p = 0.246) (see Tables 3 and 4).

Table 3 Comparison of implant diameter
Table 4 Comparison of implant length

Discussion

This study was performed in order to investigate changes in primary stability within an experimental setup of different insertion protocols and insertion modes. In order to obtain a high level of diagnostic certainty, three different methods for measurement of primary stability were recorded. As a secondary outcome parameter, potential differences between implants of different length and diameter have been evaluated. Within this study, no statistically significant difference could be shown so that the groups were not separated regarding the implant’s geometry while further analyzing the insertion mode and the drilling protocol. In line with the research hypothesis, the ODP showed significant lower stability values when compared with implants inserted via SP. The optimization of loading protocols requires establishing a patient-specific protocol [14]. The relationship between shape and shear strength during implant placement modulates the bone compression [11]. Osteotomy techniques with a narrower final drill have been shown to create a better bone-to-implant contact with enhanced primary stability values [18]. As micromotions’ risk osseointegration, there is consensus that implant stability immediately and early after placement is desirable [7]. Primary stability is generally associated with the expectation of good secondary stability, which would ensure the likelihood of implant success and osseointegration under-dimensioned protocols are thought to induce a tight contact immediately after placement: in soft bone, the use of wide diameter implants and an under-dimensioned preparation has been recommended to preserve the cortical layer [4, 18]. A study by Campos et al. reported a significant increase in bone-to-implant contact between SP and under-dimensioned protocols, but they could not show any effect in histometrical parameters [7]. Okazaki et al. analyzed the impact of different loading protocols on implant removal torque values and showed stability values to be up to eleven times lower at the time of insertion in ODPs, whereas no difference between SP and ODP could be observed after a healing phase of 6 to 12 weeks [10]. However, caution is recommended when using under-dimensioned drilling protocols: although high insertion torques ensure a greater initial implant stability and prevent adverse micromotions under loading, the induced over-compression could jeopardize the healing process [10, 28]. In addition, high stress is known to alter angiogenesis and impair new vessel formations, to induce local hypoxia and necrosis, inhibiting new bone formation and to adversely affect implant stability [4, 7, 10, 28, 37]. Also, higher degrees of primary stability do not necessarily translate into high degrees of secondary stability [11]. After a latency of 1 week, necrotic bone fills the space between bone and implant whereupon a bone remodeling takes place within 1 to 3 weeks [11, 37]. By using larger drilling dimensions, the bone-implant interface is filled with a blood clot so intermembranous-like ossification takes place without a formation of necrotic bone spots [7]. With a shorter healing period of two instead of 4 weeks, this “healing chamber” is considered a key factor for secondary stability. Due to its different healing patterns, it does not have to undergo tissue remodeling [11, 38, 39]. Contrary to this, ODPs could lead to displacements above 50–150 μm resulting in fibrous bone formation and a poor long-term stability [14]. With a reduction in primary stability and a secondary stability similar to SPs [10], test results have shown a shorter healing period [11, 38, 39] with lower degrees of compressive stress transferred to the surrounding bone. Nevertheless, in this study, implants inserted via ODP exceeded the clinically established lower limits for primary insertion. A minimum of 45 Ncm in IT and 70 units in ISQ was reported to be required to obtain the necessary stability for primary insertion [40]. Therefore, based on the results at hand, the use of ODPs might be preferable to reduce the level of compressive stress, especially in patients with dense bone.

Contrary to the research hypothesis, there was no difference in primary stability between manually and machine-driven inserted implants. To date, little is known about the influence of the insertion mode on the dental implant primary stability. Novsak et al. assumed a better primary stability in implants inserted manually and suspected that this behavior was related to a higher tactile sensation in hand-driven implantation [26]. In contrast, movements during insertion and a greater amount of rotational stress due to higher torque values could explain a poor outcome during manual insertion. In accordance with the results of the present study, it can be assumed that there is no difference between manual and machine-driven insertion modes and that surgical experience in general and especially in using either procedure seems to be more important than the insertion mode itself.

Conclusions

The aim of this study was to analyze the influence of protocol and insertion mode on dental implant primary stability ex vivo. Implants inserted via standard protocol showed higher stability values than implants inserted via an over-dimensioned protocol. Interestingly, the study demonstrated that the latter nevertheless exceeded the primary stability recommended for immediate implant insertion so that the use of ODPs might be preferable to reduce the level of compressive stress. This study also showed that manual and machine-driven insertion modes exhibited equivalent primary stability values. Under the limitations of this study, it can be hypothesized that the choice of implant type in combination with the surgical drilling technique adapted to the quality of the bony implant site [1, 2, 4, 6,7,8] as well as patient-related factors [15] and the surgical experience [8] seems to be more important than the insertion mode itself. In the absence of guidelines and in consideration of the high number of implants inserted each year, further studies are needed to confirm or disprove these results and to improve the knowledge within this area.

Availability of data and materials

The dataset supporting the conclusions of the article is included within the article and its additional files. The raw data analyzed during the current study is available from the corresponding author on reasonable request.

Abbreviations

DIS:

Dental implant primary stability

ISQ:

Implant stability quotient

IT:

Insertion torque

ODP:

Over-dimensioned protocol

RFA:

Resonance frequency analysis

SP:

Standard protocol

TO:

Torque out

References

  1. Falco A, Berardini M, Trisi P. Correlation between implant geometry, implant surface, insertion torque, and primary stability: in vitro biomechanical analysis. Int J Oral Maxillofac Implants. 2018;33(4):824–30.

    Article  Google Scholar 

  2. Lozano-Carrascal N, Salomo-Coll O, Gilabert-Cerda M, Farre-Pages N, Gargallo-Albiol J, Hernandez-Alfaro F. Effect of implant macro-design on primary stability: a prospective clinical study. Med Oral Patol Oral Cir Bucal. 2016;21(2):e214–21.

    Article  Google Scholar 

  3. Queiroz TP, Souza FA, Okamoto R, Luvizuto ER, Margonar R, Garcia IR Jr, et al. Compromised primary implant stability. J Craniofac Surg. 2012;23(5):e524–8.

    Article  Google Scholar 

  4. Khayat PG, Arnal HM, Tourbah BI, Sennerby L. Clinical outcome of dental implants placed with high insertion torques (up to 176 Ncm). Clin Implant Dent Relat Res. 2013;15(2):227–33.

    Article  Google Scholar 

  5. Degidi M, Daprile G, Piattelli A, Carinci F. Evaluation of factors influencing resonance frequency analysis values, at insertion surgery, of implants placed in sinus-augmented and nongrafted sites. Clin Implant Dent Relat Res. 2007;9(3):144–9.

    Article  Google Scholar 

  6. Gomez-Polo M, Ortega R, Gomez-Polo C, Martin C, Celemin A, Del Rio J. Does length, diameter, or bone quality affect primary and secondary stability in self-tapping dental implants? J Oral Maxillofac Surg. 2016;74(7):1344–53.

    Article  Google Scholar 

  7. Campos FE, Gomes JB, Marin C, Teixeira HS, Suzuki M, Witek L, et al. Effect of drilling dimension on implant placement torque and early osseointegration stages: an experimental study in dogs. J Oral Maxillofac Surg. 2012;70(1):e43–50.

    Article  Google Scholar 

  8. Maiorana C, Farronato D, Pieroni S, Cicciu M, Andreoni D, Santoro F. A four-year survival rate multicenter prospective clinical study on 377 implants: correlations between implant insertion torque, diameter, and bone quality. J Oral Implantol. 2015;41(3):e60–5.

    Article  Google Scholar 

  9. Swami V, Vijayaraghavan V, Swami V. Current trends to measure implant stability. J Indian Prosthodont Soc. 2016;16(2):124–30.

    Article  Google Scholar 

  10. Okazaki J, Komasa Y, Sakai D, Kamada A, Ikeo T, Toda I, et al. A torque removal study on the primary stability of orthodontic titanium screw mini-implants in the cortical bone of dog femurs. Int J Oral Maxillofac Surg. 2008;37(7):647–50.

    Article  Google Scholar 

  11. do Carmo Filho LC, Faot F, de Matos Madruga M, Marcello-Machado RM, Bordin D, Del Bel Cury AA. Effect of implant macrogeometry on peri-implant healing outcomes: a randomized clinical trial. Clin Oral Investig. 2018.

  12. Makary C, Rebaudi A, Sammartino G, Naaman N. Implant primary stability determined by resonance frequency analysis: correlation with insertion torque, histologic bone volume, and torsional stability at 6 weeks. Implant Dent. 2012;21(6):474–80.

    Article  Google Scholar 

  13. Foundation for Oral Rehabilitation. Treatment guidelines: implant insertion torque 2019 [Available from: https://www.for.org/en/treat/treatment-guidelines/single-tooth/treatment-procedures/treatment-procedures-general-considerations/implant-insertion-torque.

  14. Zanetti EM, Pascoletti G, Cali M, Bignardi C, Franceschini G. Clinical assessment of dental implant stability during follow-up: what is actually measured, and perspectives. Biosensors (Basel). 2018;8(3).

  15. Mesa F, Munoz R, Noguerol B, de Dios LJ, Galindo P, O'Valle F. Multivariate study of factors influencing primary dental implant stability. Clin Oral Implants Res. 2008;19(2):196–200.

    Article  Google Scholar 

  16. Huang H, Wismeijer D, Shao X, Wu G. Mathematical evaluation of the influence of multiple factors on implant stability quotient values in clinical practice: a retrospective study. Ther Clin Risk Manag. 2016;12:1525–32.

    Article  Google Scholar 

  17. Osstell. The Technique Behind Osstell 2019 [Available from: https://www.osstell.com/clinical-guidelines/the-technique-behind-osstell/.

  18. Bilhan H, Geckili O, Mumcu E, Bozdag E, Sunbuloglu E, Kutay O. Influence of surgical technique, implant shape and diameter on the primary stability in cancellous bone. J Oral Rehabil. 2010;37(12):900–7.

    Article  Google Scholar 

  19. Zix J, Kessler-Liechti G, Mericske-Stern R. Stability measurements of 1-stage implants in the maxilla by means of resonance frequency analysis: a pilot study. Int J Oral Maxillofac Implants. 2005;20(5):747–52.

    PubMed  Google Scholar 

  20. Rabel A, Kohler SG, Schmidt-Westhausen AM. Clinical study on the primary stability of two dental implant systems with resonance frequency analysis. Clin Oral Investig. 2007;11(3):257–65.

    Article  Google Scholar 

  21. Akkocaoglu M, Uysal S, Tekdemir I, Akca K, Cehreli MC. Implant design and intraosseous stability of immediately placed implants: a human cadaver study. Clin Oral Implants Res. 2005;16(2):202–9.

    Article  Google Scholar 

  22. Carlsson L, Rostlund T, Albrektsson B, Albrektsson T. Removal torques for polished and rough titanium implants. Int J Oral Maxillofac Implants. 1988;3(1):21–4.

    PubMed  Google Scholar 

  23. Motoyoshi M, Hirabayashi M, Uemura M, Shimizu N. Recommended placement torque when tightening an orthodontic mini-implant. Clin Oral Implants Res. 2006;17(1):109–14.

    Article  Google Scholar 

  24. Simon H, Caputo AA. Removal torque of immediately loaded transitional endosseous implants in human subjects. Int J Oral Maxillofac Implants. 2002;17(6):839–45.

    PubMed  Google Scholar 

  25. Wennerberg A, Albrektsson T, Lausmaa J. Torque and histomorphometric evaluation of c.p. titanium screws blasted with 25- and 75-microns-sized particles of Al2O3. J Biomed Mater Res. 1996;30(2):251–60.

    Article  Google Scholar 

  26. Novsak D, Trinajstic Zrinski M, Spalj S. Machine-driven versus manual insertion mode: influence on primary stability of orthodontic mini-implants. Implant Dent. 2015;24(1):31–6.

    PubMed  Google Scholar 

  27. Ziebart J, Fan S, Schulze C, Kammerer PW, Bader R, Jonitz-Heincke A. Effects of interfacial micromotions on vitality and differentiation of human osteoblasts. Bone Joint Res. 2018;7(2):187–95.

    Article  Google Scholar 

  28. Griggs JA. Dental implants. Dent Clin North Am. 2017;61(4):857–71.

    Article  Google Scholar 

  29. Blanco J, Alvarez E, Munoz F, Linares A, Cantalapiedra A. Influence on early osseointegration of dental implants installed with two different drilling protocols: a histomorphometric study in rabbit. Clin Oral Implants Res. 2011;22(1):92–9.

    Article  Google Scholar 

  30. Kim MS, Jung UW, Kim S, Lee JS, Lee IS, Choi SH. Bone apposition on implants coated with calcium phosphate by ion beam assisted deposition in oversized drilled sockets: a histologic and histometric analysis in dogs. J Periodontal Implant Sci. 2013;43(1):18–23.

    Article  Google Scholar 

  31. Campos FE, Jimbo R, Bonfante EA, Barbosa DZ, Oliveira MT, Janal MN, et al. Are insertion torque and early osseointegration proportional? A histologic evaluation. Clin Oral Implants Res. 2015;26(11):1256–60.

    Article  Google Scholar 

  32. Huang HM. Chee TJ. Feng SW. Modified surgical drilling protocols influence osseointegration performance and predict value of implant stability parameters during implant healing process. Clin Oral Investig: Lew WZ; 2020.

    Google Scholar 

  33. Hi-Tec Implants. Tapered Self Thread - Internal Hex Thread Implants 2019 [Available from: http://hitec-implants.com/Implant1.html.

  34. Hi-Tec Implants LTD. Product Catalogue; 2015.

    Google Scholar 

  35. Bolm I, Goetze E, Kammerer PW, Sader R, Klos M, Landes C, et al. Self-drilling and self-tapping miniscrews for osteosynthesis fixture after LeFort I osteotomy: an ex vivo trial for primary stability and a randomized clinical study. J Surg Res. 2017;212:246–52.

    Article  Google Scholar 

  36. Cohen J. Statistical power analysis for the behavioral sciences. 2 ed: LAWRENCE ERLBAUM ASSOCIATES, PUBLISHERS 1988.

  37. Goswami MM, Kumar M, Vats A, Bansal AS. Evaluation of dental implant insertion torque using a manual ratchet. Med J Armed Forces India. 2015;71(Suppl 2):S327–32.

    Article  Google Scholar 

  38. Marin C, Bonfante E, Granato R, Neiva R, Gil LF, Marao HF, et al. The effect of osteotomy dimension on implant insertion torque, healing mode, and osseointegration indicators: a study in dogs. implant dent. 2016;25(6):739-43.

  39. Baires-Campos FE, Jimbo R, Bonfante EA, Fonseca-Oliveira MT, Moura C, Zanetta-Barbosa D, et al. Drilling dimension effects in early stages of osseointegration and implant stability in a canine model. Med Oral Patol Oral Cir Bucal. 2015;20(4):e471–9.

    Article  Google Scholar 

  40. Lages FS, Douglas-de Oliveira DW, Costa FO. Relationship between implant stability measurements obtained by insertion torque and resonance frequency analysis: a systematic review. Clin Implant Dent Relat Res. 2018;20(1):26–33.

    Article  Google Scholar 

Download references

Acknowledgements

Not applicable

Funding

No funding.

Author information

Authors and Affiliations

Authors

Contributions

Implant insertion and data collection were performed by DH, PK, and EG. HS, PK, EG, and BA contributed to research design. Data analysis, statistic and interpretation, as well as drafting the article were performed by HS, PK, EG, BA, and DH. DH AND PK were major contributors in writing the manuscript. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Diana Heimes.

Ethics declarations

Ethics approval and consent to participate

Not applicable

Consent for publication

Not applicable

Competing interests

Henning Staedt, Peer W. Kämmerer, Elisabeth Goetze, Daniel G. E. Thiem, Bilal Al-Nawas, and Diana Heimes declare that they have no competing interests.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Staedt, H., Kämmerer, P.W., Goetze, E. et al. Implant primary stability depending on protocol and insertion mode — an ex vivo study. Int J Implant Dent 6, 49 (2020). https://doi.org/10.1186/s40729-020-00245-3

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s40729-020-00245-3

Keywords