Influence of exposure of customized dental implant abutments to different cleaning procedures: an in vitro study using AI-assisted SEM/EDS analysis

Abstract

Purpose

Dental implant abutments are defined as medical devices by their intended use. Surfaces of custom-made CAD/CAM two-piece abutments may become contaminated during the manufacturing process in the dental lab. Inadequate reprocessing prior to patient care may contribute to implant-associated complications. Risk-adapted hygiene management is required to meet the requirements for medical devices.

Methods

A total of 49 CAD/CAM-manufactured zirconia copings were bonded to prefabricated titanium bases. One group was bonded, polished, and cleaned separately in dental laboratories throughout Germany (LA). Another group was left untreated (NC). Five groups received the following hygiene regimen: three-stage ultrasonic cleaning (CP and FP), steam (SC), argon–oxygen plasma (PL), and simple ultrasonic cleaning (UD). Contaminants were detected using scanning electron microscopy (SEM) and energy-dispersive X-ray spectroscopy (EDS) and segmented and quantified using interactive machine learning (ML) and thresholding (SW). The data were statistically analysed using non-parametric tests (Kruskal–Wallis test, Dunn’s test).

Results

Significant differences in contamination levels with the different cleaning procedures were found (p ≤ 0.01). The FP–NC/LA groups showed the most significant difference in contamination levels for both measurement methods (ML, SW), followed by CP–LA/NC and UD–LA/NC for SW and CP–LA/NC and PL–LA/NC for ML (p ≤ 0.05). EDS revealed organic contamination in all specimens; traces of aluminum, silicon, and calcium were detected.

Conclusions

Chemothermal cleaning methods based on ultrasound and argon–oxygen plasma effectively removed process-related contamination from zirconia surfaces. Machine learning is a promising assessment tool for quantifying and monitoring external contamination on zirconia abutments.

Graphical Abstract

Background

Implant abutments in the dental implant system are the connecting element between implant-supported prostheses/crowns and endosseous implant bodies. This titanium or all-ceramic foreign body remains permanently in direct contact with the peri-implant mucosa in a critical and bacteria-rich environment between the peri-implant bone and the oral cavity as a transgingival support for the artificial tooth or implant-supported prosthesis. As the barrier function of the peri-implant soft tissue collar is inferior to that of natural teeth, recent research has increasingly focused on improving the quality of the peri-implant soft tissue interface, especially regarding cleaning procedures before placement [1,2,3,4,5,6,7,8,9,10,11].

To address the requirements of aesthetically sensitive restoration areas, customised all-ceramic CAD/CAM abutments are increasingly used to achieve an anatomically correct emergence profile. Two-piece implant abutments consisting of a CAD/CAM-manufactured zirconia coping bonded to a premanufactured titanium base are currently the first choice for demanding restorations. This is due to their excellent biocompatibility, the improved material properties with precise fit at the implant–abutment interface, and the improved tooth-coloured masking of the underlying titanium implant [8, 12,13,14,15,16]. Nowadays, CAD/CAM production is carried out in a central milling facility, a dental lab, or with an in-office milling machine [17]. However, implant superstructures can become contaminated during manufacturing, transport, and packaging [18, 19]. In particular, customized two-piece abutments have shown higher contamination levels after reprocessing than one-piece abutments [2, 5]. According to several laws, regulations and recommendations, implant abutments are classified by manufacturers as medical devices based on their intended use [20,21,22]. Dental practitioners use custom-made medical devices on a daily basis. They need to be aware of the risk-adapted classification to perform appropriate hygiene management before implant abutment placement. As defined in the European Medical Devices Regulation (MDR EU 2017/745), dental implant abutments are:

  • intended for long-term use

  • an invasive device

  • an implantable device

Such devices are classified according to the MDR classification rules as follows: “class IIb if they are intended for long-term use, except if they are used in the oral cavity as far as the pharynx, in an ear canal up to the ear drum or in the nasal cavity and are not liable to be absorbed by the mucous membrane, in which case they are classified as class IIa” [20]. This classification is also consistent with the FDA’s medical device regulations [Code of Federal Regulations (CFR): 21 CFR 872.3630] [23].

Risk-adapted reprocessing (cleaning, disinfection, and, if required, sterilisation) can be derived from medical device classification [21, 22]. To this end, dentists now have unprecedented reprocessing strategies with appropriate substances and instruments [24]. However, the cleaning efficacy of cleaning and disinfection procedures for custom-made implant prosthetic components has yet to be sufficiently scientifically investigated. This is due in part to the lack of tools to quantitatively measure and monitor process-related contamination before and after different cleaning procedures and in part to the fact that, according to the current state of research, there is a lack of knowledge about the exact clinical risks of contaminated implantable devices. Binding limit values for acceptable contamination of medical devices still need to be added [25, 26].

This study aimed to determine the level of contamination on zirconia surfaces at the mucosal transition zone between the oral cavity and peri-implant bone after fabrication and reprocessing. Second, identify suitable cleaning and disinfection procedures for reprocessing customized all-ceramic abutments. The null hypothesis to be tested and confirmed was that the mean values did not differ significantly between the different cleaning methods.

Methods

Sample preparation and cleaning procedures

In this study, 49 customised two-piece implant abutments were fabricated in a dental laboratory. Based on a patient case for the implant-retained single crown in FDI position 14, an all-ceramic CAD/CAM zirconia coping (Anatomic coloured A2, Zirkonzahn, Gais, Italy) was bonded to a prefabricated titanium base (Titanium base CAD/CAM, 4.3 mm diameter, CAMLOG Biotechnologies AG, Basel, Switzerland). The external geometry had a prosthetic height of 8.20 mm and an abutment shoulder width of 5.70 mm. The two-piece zirconia abutments were randomly divided into seven study groups of seven specimens each (Fig. 1). In group LA, an unopened premanufactured titanium base, an already milled zirconia abutment, and the resin cement (Multilink Hybrid Abutment, Ivoclar Vivadent, Schaan, Liechtenstein) were sent to seven dental laboratories throughout Germany with the instruction to return the two-piece abutments cleaned according to their protocol and ready for clinical use. It was unclear how the laboratories performed cleaning and disinfection. The bonding surfaces of the 42 remaining two-piece titanium base and all-ceramic copings were blasted with aluminum oxide particles (Al2O3) 50 µm in size (Cobra 50 µm, Renfert, Hilzingen, Germany) at a reduced pressure of 0.8 bar and a distance of 10 mm in the dental laboratory (spot blasting unit P-G 400, Hanisch + Rieth, Winterbach, Germany) [27]. The surfaces were then steamed for 30 s and dried with oil-free air. The bonding surfaces were marked with a permanent marker before sandblasting [28]. The surfaces were sandblasted until the colour was removed entirely. The titanium base and zirconia abutment cleaned bonding surfaces were conditioned (Monobond plus, Ivoclar Vivadent, Schaan, Liechtenstein) and bonded with resin cement according to the manufacturer's instructions (Multilink Hybrid Abutment, Ivoclar Vivadent). Subsequently, the excess cement was removed, and the specimens were polished in two steps using ceramic polishers with diamond grit (94003C and 94003M, Gebr. Brasseler/Komet Dental, Lemgo, Germany) [29]. After the final surface treatment, specimens were randomly grouped into seven pieces and subjected to six different hygiene regimes: no cleaning (NC), multi-stage ultrasonic cleaning with various disinfectants (CP, FP), steam cleaning (SC), low-pressure argon–oxygen plasma cleaning (PL) and single-stage ultrasonic cleaning and disinfection (UD) (full description in Table 1). The hygiene management in the laboratory-based group (LA) was not subsequently queried and remains unclear. After cleaning, all specimens were individually sealed in sterilisation pouches (HS-Sterifoil, Henry Schein Dental, Langen, Germany) (Fig. 2). The materials used in this study are listed in Table 2.

Fig. 1
figure 1

Study design shows the number of cleaning groups and the evaluation process. Created with BioRender.com

Table 1 Cleaning and disinfection methods used in this study
Table 2 Materials used in this study

Imaging acquisition (SEM) and image data processing and analysis

All specimens were subjected to SEM and EDS analysis (Phenom-World B.V., Eindhoven, The Netherlands). The scanning electron microscope is equipped with a highly sensitive backscattered electron (BSE) detector and operates at an accelerating voltage of 15 kV. Backscattered electron imaging allows conclusions to be drawn about the chemical nature and location of various contaminants on the sample. Low atomic number elements, such as carbon, appear relatively dark, while higher atomic number elements, such as zirconium, appear relatively bright. The high spatial resolution and large field of view of a stitched image allowed us to filter out areas of interest (contaminations) for subsequent EDS analysis. Two open-source programs were used to classify, segment, and quantify the contaminants: Fiji (ImageJ, version 1.53c) and ilastik (Ilastik, version 1.3.3) [30, 31]. Two in-house developed workflows based on pixel-based machine learning (ML) and thresholding (SW) were compared. A detailed image processing and analysis steps description has already been published [25].

Semiquantitative chemical surface analysis (EDS)

All specimens were evaluated with EDS point analyses to determine the chemical composition of the contaminations in the predefined observation field. Due to the fully integrated thermoelectrically cooled silicon drift detector (SDD), all measurements were performed in temporal relation to the SEM analysis. At least two different points for each specimen were analysed for elemental composition at 2500 × magnification. The working distance was 6 mm with a field of view (FOV) of 108 µm. Excited by the interactions between the irradiated electrons and the electromagnetic field of the investigated samples, the characteristic X-ray radiation was emitted element-specifically. The accelerating voltage was 15 kV, and the active detector had an area of 25 mm2 with a take-off angle (TOA) of 29°. The peak position identifies the element, while the intensity of the signal indicates the number of X-ray quanta detected per element. An automatic peak deconvolution algorithm evaluated the spectrum through the coupled software (Phenom Elemental Identification Version 3.8.4.0, ThermoFisher Scientific, Eindhoven, The Netherlands). For peak identification of the acquired spectrum, we deactivated elements with a lower match below a peak match of ≥ 0.95. The relative concentration fractions of the detected elements (in atomic per cent) were calculated for all groups. Elements ≤ 1 atomic per cent (at. %) were filtered out (Table 3).

Statistical analysis

Data were analysed using a statistical program (SPSS version 27.0, IBM SPSS). Nonparametric methods (Kruskal–Wallis test) were used to test the influence of the cleaning method on the level of contamination. Dunn’s test was used for multiple comparisons between cleaning methods and contamination levels. p values for subsequent multiple comparisons were corrected according to the Bonferroni method. The effect size was measured using Cohen’s d. Descriptive data analysis was used to quantitatively assess contamination levels and EDS analysis. All statistical tests were performed at a two-sided significance level of α = 0.05 (Fig. 2).

Fig. 2
figure 2

CAD/CAM manufacturing, reprocessing, and insertion of dental implant abutments. Created with BioRender.com

Results

A summary of the data and statistical analyses is shown in Figs. 3, and 4, Tables 3, 4 and 5.

Quantification using interactive machine learning

The contamination level of specimens cleaned by the FP (minimum 0.0018%; median 0.0025%; maximum 0.0084%), CP (minimum 0.0005%; median 0.0064%; maximum 0.0140%) and PL (minimum 0.0011%; median 0.0044%; maximum 0. 0093%) showed the lowest residual contamination compared to the UD (minimum 0.0042%; median 0.0069%; maximum 0.0367%), SC (minimum 0.0045%; median 0.0077%; maximum 0.0437%), NC (minimum 0.0380%; median 0.0692%; maximum 0.0886%) and LA (minimum 0.0216%; median 0.0995%; maximum 0.3089%) methods (Fig. 3). A Kruskal–Wallis test indicated significant differences between contamination levels and cleaning methods [H(6) = 33.8, p ≤ 0.00001]. Subsequent post hoc tests (Dunn–Bonferroni tests) showed the greatest difference in contamination levels for the FP–LA/NC groups, followed by CP/LA, PL/LA, CP/NC and PL/NC (p ≤ 0.05, UD/LA, UD/NC, SC/LA, FP/SC, SC/NC, FP–CP/PL/UD, CP–PL/UD/SC, PL–UD/SC, UD/SC, NC/LA were not significantly different) (Table 3). All significant group comparisons had a large effect size according to Cohen’s d. The most significant effect was seen in the FP/NC, PL/NC and CP/NC group comparisons with d = 4.92, d = 4.75 and d = 4.60, respectively (Fig. 3, Table 3).

Fig. 3
figure 3

Box-plot diagram of the contamination level in per cent for ML. One outlier of 0.31% in the LA group has been removed for clarity

Table 3 Statistical analysis of the cleaning method’s influence on the cleaning efficiency for method ML

Threshold-based quantification

The contamination level of specimens cleaned by the CP (minimum 0.0003%; median 0.0098%; maximum 0.0132%), FP (minimum 0.0004%; median 0.0015%; maximum 0.0221%) and UD (minimum 0.0016%; median 0.0074%; maximum 0. 0.0233%) was lower than for specimens cleaned by the SC (minimum 0.0024%; median 0.0097%; maximum 0.0355%), PL (minimum 0.0003%; median 0.0109%; maximum 0.0134%), NC (minimum 0.0305%; median 0.0356%; maximum 0.0652%) and LA (minimum 0.0135%; median 0.0730%; maximum 0.2895%) methods (Fig. 4). A Kruskal–Wallis test showed that the level of contamination was influenced by the cleaning methods [H(6) = 28.2, p ≤ 0.0001]. Subsequent post hoc tests (Dunn–Bonferroni tests) showed the best decontamination for the FP method, followed by CP and UD (p ≤ 0.05, FP–CP/UD/PL/SC, CP–UD/PL/SC, UD–PL/SC, PL–SC/NC/LA, SC–NC/LA and NC/LA were not significantly different) (Table 4). All significant group comparisons had a large effect size according to Cohen’s d. The CP/NC and FP/NC group differences had the most significant effect size with d = 3.36 and d = 3.00, respectively (Fig. 4, Tables 4, 5).

Fig. 4
figure 4

Box-plot diagram of the contamination level in per cent for SW. One outlier of 0.29% in the LA group has been removed for clarity

Table 4 Statistical analysis of the cleaning method’s influence on the cleaning efficiency for method SW
Table 5 Descriptive statistics for comparison of measurement methods in per cent

EDS analysis

The elemental analysis of all cleaning groups by EDS point analysis is summarized in Table 6. The presence of four major elements was detected in all observation areas of all cleaning groups: carbon (C), oxygen (O), zirconium (Zr) and yttrium (Y), the latter three being associated with the base material of the samples. Traces of aluminum, silicon and iron were also detected, with a maximum value of iron of 33.3 at. %, followed by aluminum at 25.9 at. % and silicon at 17.5 at. %. Differential EDS was used to separate the received X-ray quanta of the surface contaminants from the signal of the subsurface material. Therefore, the signals of the uncontaminated abutment material were subtracted from the mixed signal of the surface contamination. A representative elemental analysis of contamination spots using SEM–BSE is shown in Fig. 5. These included a contaminant spot from the NC and LA groups (full results in Table 6). The subtracted spectrum shows an organic composition of two major elements: carbon (C) and oxygen (O), with values above 10 at. % (Fig. 5A–F) and an inorganic composition consisting of a metal alloy rich in cobalt (Co) and chromium (Cr) (Fig. 5G–L). To visually check for alterations of the interfaces between the zirconia coping and the titanium base after hygiene management, representative scanning electron micrographs of the adhesive joint from all study groups were presented at 500 × magnification (Fig. 6).

Table 6 Descriptive statistics of the EDS analysis

Discussion

Implant abutments are in direct contact with the peri-implant mucosa. The medical device user (dentist) is responsible for risk management before reprocessing and patient care (Fig. 2). Decision-making depends on the type of application; the dentist must classify customised CAD/CAM-manufactured abutments as risk-adapted, considering the manufacturer’s specifications (DIN EN ISO 17664) [21, 22]. As customised implant abutments are often used for late implant-prosthetic rehabilitation after wound healing is complete, a semi-critical classification is usually sufficient [20]. If there is doubt about the classification, the more critical classification should be used [20, 21]. Hygiene management for semi-critical devices must include validated cleaning and disinfection and, if classified as critical, sterilisation prior to patient care. Dentists must also consider appropriate packaging systems and device transportation, as recontaminating the product is possible [18, 20]. However, following the manufacturing process in the dental laboratory, contaminants such as carbon, titanium and aluminum microparticles have been found on the abutment surface adjacent to the mucosal transition zone, even after reprocessing [2, 5, 18]. In addition, customised implant abutments appear to have higher levels of contamination than prefabricated ones [2, 5]. To date, no cleaning and disinfection method has completely removed process-related contamination. The issue of valid reprocessing of implant abutments continues to be debated [32,33,34,35]. However, the method for monitoring and measuring decontamination has yet to be researched [25]. When reviewing the literature, no established “gold standard” reference method exists. To date, only a few studies have been identified that have quantified contamination on implant abutments. Contamination detection has so far relied on manual annotation of SEM images by experts [2, 5, 19, 36].

In the present in vitro study, interactive machine learning was used for the first time for classification and segmentation compared to segmentation by thresholding. In addition, quantification was performed in a simple self-written program [25]. Not all hygiene regimes could reduce contamination levels significantly—there were significant cleaning-specific differences. The highest surface cleanliness was achieved by chemothermal ultrasonic cleaning (CP, FP and UD) and argon–oxygen plasma treatment (PL) (Tables 3, 4). The most significant difference in the contamination level between the FP/NC, CP/NC, UD/NC and PL/NC groups should be highlighted. The results are consistent with Canullo et al. and Gehrke et al., who also reported superior ultrasonic cleaning for residual and bacterial decontamination [2, 5, 37]. The most effective surface cleanliness for ultrasonic cleaning was measured by Canullo et al. at 0.0007%; generally higher contamination levels were measured for two-piece abutments as detected by Gehrke et al., even with multi-stage ultrasonic cleaning, which is also consistent with the current measurements [2, 5]. Nevertheless, the data from this study showed deviations from the measured values of Canullo et al. and Gehrke et al. This could be due to differences in material selection, analysis tools, image data processing, and evaluation. While ultrasonic reprocessing with chemical disinfectants follows national and international regulations, plasma treatment is not currently a validated reprocessing strategy [20, 21, 33]. However, its application to all-ceramic biomaterials has been described [4, 38, 39]. Plasma treatment of zirconia resulted in comparable cell adhesion to ultrasound and disinfection, increased collagen fiber density and stable peri-implant bone level compared to steam cleaning measured on titanium abutments [3, 4, 38]. In addition, the cleaning and disinfecting effect has already been highlighted, with no adverse impact on tensile bond strength [2, 40, 41]. However, the data presented here contradicts the efficiency of plasma cleaning published by Farronato et al. His measurements showed that the use of argon plasma alone was inferior to that of argon plasma in combination with ultrasonic cleaning [36]. Against this contradiction, it can be argued that image segmentation was done by visual grid assessment. Squares containing contaminants were scored positive and compared as a percentage of the total number of squares in the grid. This may have led to an inaccuracy compared to direct contamination detection by supervised machine learning or thresholding [25]. Uncleaned specimens showed similar contamination levels to previous studies of one-piece titanium and two-piece zirconia abutments [5, 19]. Steam cleaning, although commonly used clinically, is not a validated reprocessing method, as discussed by Kern et al. [32]. Although Canullo et al. had shown that thermal cleaning with steam for 30 s reduced residual contamination and a cleaning effect was also observed in the present in vitro study, it was significantly different from cleaning with ultrasound and argon oxygen plasma [19] (Tables 3, 4). This was also reflected at the cellular level in lower cell viability and increased bacterial colonization after steam cleaning compared to ultrasound reprocessing [11, 42]. The LA group revealed a wide range of data with reduced efficacy of final cleaning and disinfection methods. Although the data are limited and should be interpreted with caution due to the small sample size, they illustrate the heterogeneity of approaches to hygiene management between dentists and dental laboratories in Germany. This suggests that adequate reprocessing of semi-critical and critical medical devices, such as implant abutments, needs to be adequately implemented.

While the agreement with previous manual expert annotations supports the reliability of the new AI-assisted detection and quantification method, robust validation on independent data and larger heterogeneous data sets currently needs to be improved [25, 43].

As EDS is only an auxiliary tool for morphological characterisation, the results of EDS analysis must be interpreted with limitations. EDS analysis cannot detect carbonaceous contaminants in the molecular range (e.g., from ambient air or packaging) [44]. However, carbonaceous particles in the micrometre range could be identified, which explains the strong carbon signal in almost all measurements (Table 6). On average, the overlying contaminants consisted of 40.5 at. % carbon and metallic contaminants such as aluminum and titanium were detected in isolated cases (Fig. 5). This is agreed with previous spectroscopic investigations [5, 18, 19, 42]. The EDS point analysis is limited in its quantitative significance, and assigning the results to individual cleaning methods was impossible.

Fig. 5
figure 5

Examples of the elemental composition of the different contaminants and the background composition (zirconia) with Phenom ProX–SEM–EDS; measuring points are marked with a cross and number, percentages in atomic per cent: A–F sample from group NC, B shows the EDS spectrum for spot analysis spot 1, D shows the EDS spectrum for spot analysis spot 2, E, F subtracted spectra of the sample from group NC: spectrum 1–spectrum 2; G–L sample from group LA, H shows the EDS spectrum for spot analysis spot 1, J shows the EDS spectrum for spot analysis spot 2, K, L subtracted spectra for the sample from group LA: spectrum 1–spectrum 2

SEM analysis of the adhesive joint of the two-piece specimens revealed irregular pits and inclusions that altered the circular milling grooves. In addition, wear particles were detected qualitatively on all prefabricated titanium bases (Fig. 6). This surface roughness and contamination may be attributed to the final manufacturing processes (airborne particle abrasion (APA), bonding, polishing). Different surface cleanliness levels could be observed visually depending on the cleaning method. The specimens exposed to the CP cleaning protocol with acetone-containing solution showed partial dissolution of the adhesive joint (Fig. 6B). While previous studies on pure titanium surfaces have shown no differences in surface properties, adverse effects on the tensile bond strength of two-piece abutments have not been investigated [45]. Further in vitro studies should follow to clarify this issue, but in the meantime, the use of acetone-containing cleaning solutions for hybrid structures should be avoided.

Fig. 6
figure 6

Scanning electron micrographs of the adhesive joint after various cleaning and disinfection procedures (500 × magnification, 537 µm field of view), white arrows mark the adhesive joint: A camera overview image, square area selection at the interface between zirconia abutment, adhesive joint, and titanium base, B group CP, C group FP, D group NC, E group SC, F group PL, G group LA, H group UD

While the effects of surface topography and chemical composition of zirconia on the peri-implant soft tissue are well-understood and make clinical treatment more predictable, the potential negative impact of residual contamination on zirconia abutments in the implant–abutment interface has not been sufficiently investigated [42, 46,47,48,49,50]. In particular, there is a lack of precise knowledge about possible thresholds for “acceptable levels of contamination” of medical devices and in vivo data on immunological responses to contaminated implanted biomaterials that may disrupt the peri-implant tissue seal over time and promote marginal bone resorption. However, contaminations and remnants of handling and manufacture identified on abutments in this study must be related to the concept of foreign body equilibrium, which Albrektsson et al. established years ago to explain marginal bone loss and disturbed osseointegration [51, 52]. An imbalance of this patient-specific foreign body equilibrium, e.g., by the additional entry of foreign microparticles into the transition zone between soft tissue and bone, could lead to immune-mediated foreign body reactions up to peri-implantitis and bone loss. Further studies are required to investigate the relationship between possible individual input of inflammatory immune processes due to process-related residual contamination in the mucosal transition zone and adverse events in the peri-implant tissue. Considering the practical implications of international hygiene regulations, it can be summarised that the user (dentist) has a duty of care in managing invasive medical devices, e.g., customised implant abutments. In this regard, medical devices should be “manufactured in such a way as to minimise the risks posed by … particles that may be released from the device, including abrasion, degradation products and processing residues” [20]. This ensures patient safety and may improve the long-term stability of the implant-supported restoration.

Conclusions

The following conclusions can be drawn based on the data from this study.

For hygiene management of semicritical medical devices, single or multistage ultrasonic immersion baths in combination with suitable chemical disinfectants can achieve effective decontamination on zirconia surfaces. Reprocessing by plasma treatment is not currently a validated process. Although this study has shown that ultrasound-based systems can achieve comparable decontamination, future studies should verify the results. Steam cleaning is not recommended. Regarding measurement methodology, AI-assisted detection and monitoring of process-related contamination could improve manufacturers’ quality management systems, save expert resources, and overcome subjective variability.

Availability of data and materials

The data sets used and analysed during the current study are available from the corresponding author on reasonable request.

Abbreviations

SEM:

Scanning electron microscope

EDS:

Energy dispersive X-ray spectroscopy

CAD/CAM:

Computer-aided design/computer-aided manufacturing

MDR:

Medical devices regulation

CFR:

Code of Federal Regulations

BSE:

Back scattered electrons

SDD:

Silicon drift detector

FOV:

Field of view

TOA:

Take-off angle

References

  1. Canullo L. Clinical outcome study of customized zirconia abutments for single-implant restorations. Int J Prosthodont. 2007;20(5):489–93.

    PubMed  Google Scholar 

  2. Canullo L, Micarelli C, Lembo-Fazio L, Iannello G, Clementini M. Microscopical and microbiologic characterization of customized titanium abutments after different cleaning procedures. Clin Oral Implants Res. 2014;25(3):328–36.

    Article  PubMed  Google Scholar 

  3. Canullo L, Tallarico M, Penarrocha M, Corrente G, Fiorellini J, Penarrocha D. Plasma of argon cleaning treatment on implant abutments in periodontally healthy patients: six years postloading results of a randomized controlled trial. Int J Periodontics Restorative Dent. 2017;37(5):683–90.

    Article  PubMed  Google Scholar 

  4. Mehl C, Kern M, Zimmermann A, Harder S, Huth S, Selhuber-Unkel C. Impact of cleaning procedures on adhesion of living cells to three abutment materials. Int J Oral Maxillofac Implants. 2017;32(5):976–84.

    Article  PubMed  Google Scholar 

  5. Gehrke P, Abazari C, Schlichter K, Fischer C, Duddeck D, Romanos GE, et al. Qualitative and semi-quantitative assessment of processing-related surface contamination of one- and two-piece CAD/CAM abutments before and after ultrasonic cleaning. Materials (Basel). 2020;13(14):3225.

    Article  PubMed  PubMed Central  Google Scholar 

  6. Sculean A, Gruber R, Bosshardt DD. Soft tissue wound healing around teeth and dental implants. J Clin Periodontol. 2014;41(Suppl 15):S6-22.

    PubMed  Google Scholar 

  7. Berglundh T, Lindhe J, Ericsson I, Marinello CP, Liljenberg B, Thomsen P. The soft tissue barrier at implants and teeth. Clin Oral Implants Res. 1991;2(2):81–90.

    Article  PubMed  Google Scholar 

  8. Abrahamsson I, Berglundh T, Glantz PO, Lindhe J. The mucosal attachment at different abutments. An experimental study in dogs. J Clin Periodontol. 1998;25(9):721–7.

    Article  PubMed  Google Scholar 

  9. Welander M, Abrahamsson I, Berglundh T. The mucosal barrier at implant abutments of different materials. Clin Oral Implants Res. 2008;19(7):635–41.

    Article  PubMed  Google Scholar 

  10. Mehl C, Kern M, Schütte AM, Kadem LF, Selhuber-Unkel C. Adhesion of living cells to abutment materials, dentin, and adhesive luting cement with different surface qualities. Dent Mater. 2016;32(12):1524–35.

    Article  PubMed  Google Scholar 

  11. Kim S, Choi C, Cha Y, Chang JS. The efficacy of convenient cleaning methods applicable for customized abutments: an in vitro study. BMC Oral Health. 2021;21(1):78.

    Article  PubMed  PubMed Central  Google Scholar 

  12. Ishikawa-Nagai S, Da Silva JD, Weber HP, Park SE. Optical phenomenon of peri-implant soft tissue. Part II. Preferred implant neck color to improve soft tissue esthetics. Clin Oral Implants Res. 2007;18(5):575–80.

    Article  PubMed  Google Scholar 

  13. Stimmelmayr M, Edelhoff D, Güth J-F, Erdelt K, Happe A, Beuer F. Wear at the titanium–titanium and the titanium–zirconia implant–abutment interface: a comparative in vitro study. Dent Mater. 2012;28(12):1215–20.

    Article  PubMed  Google Scholar 

  14. Glauser R, Sailer I, Wohlwend A, Studer S, Schibli M, Schärer P. Experimental zirconia abutments for implant-supported single-tooth restorations in esthetically demanding regions: 4-year results of a prospective clinical study. Int J Prosthodont. 2004;17(3):285–90.

    PubMed  Google Scholar 

  15. Pitta J, Zarauz C, Pjetursson B, Sailer I, Liu X, Pradies G. A systematic review and meta-analysis of the influence of abutment material on peri-implant soft tissue color measured using spectrophotometry. Int J Prosthodont. 2020;33(1):39–47.

    Article  PubMed  Google Scholar 

  16. Jung RE, Holderegger C, Sailer I, Khraisat A, Suter A, Hämmerle CH. The effect of all-ceramic and porcelain-fused-to-metal restorations on marginal peri-implant soft tissue color: a randomized controlled clinical trial. Int J Periodontics Restorative Dent. 2008;28(4):357–65.

    PubMed  Google Scholar 

  17. Beuer F, Schweiger J, Edelhoff D. Digital dentistry: an overview of recent developments for CAD/CAM generated restorations. Br Dent J. 2008;204(9):505–11.

    Article  PubMed  Google Scholar 

  18. Sawase T, Wennerberg A, Hallgren C, Albrektsson T, Baba K. Chemical and topographical surface analysis of five different implant abutments. Clin Oral Implant Res. 2000;11(1):44–50.

    Article  Google Scholar 

  19. Canullo L, Micarelli C, Iannello G. Microscopical and chemical surface characterization of the gingival portion and connection of an internal hexagon abutment before and after different technical stages of preparation. Clin Oral Implants Res. 2013;24(6):606–11.

    Article  PubMed  Google Scholar 

  20. Consolidated text: Regulation (EU) 2017/745 of the European Parliament and of the Council of 5 April 2017 on medical devices, amending Directive 2001/83/EC, Regulation (EC) No 178/2002 and Regulation (EC) No 1223/2009 and repealing Council Directives 90/385/EEC and 93/42/EEC. https://eur-lex.europa.eu/eli/reg/2017/745/2020-04-242017.

  21. Hygiene Requirements for the Reprocessing of Medical Devices. Bundesgesundheitsblatt—Gesundheitsforschung—Gesundheitsschutz. 2012;55(10):1244-310.

  22. Hygieneleitfaden. Deutscher Arbeitskreis für Hygiene in der Zahnmedizin (Hrsg). 2022;15. Ausgabe.

  23. 21 CFR 872.3630. https://www.ecfr.gov/current/title-21/chapter-I/subchapter-H/part-872/subpart-D/section-872.3630: Electronic Code of Federal Regulations (eCFR); 2004.

  24. Canullo L, Tallarico M, Chu S, Peñarrocha D, Özcan M, Pesce P. Cleaning, disinfection, and sterilization protocols employed for customized implant abutments: an international survey of 100 universities worldwide. Int J Oral Maxillofac Implants. 2017;32(4):774–8.

    Article  PubMed  Google Scholar 

  25. Hofmann P, Kunz A, Schmidt F, Beuer F, Duddeck D. Segmentation of process-related contaminations on two-piece abutments using pixel-based machine learning. A new quantification approach? Int J Comput Dent. 2023;0(0):0.

  26. Duddeck DU, Albrektsson T, Wennerberg A, Larsson C, Beuer F. On the cleanliness of different oral implant systems: a pilot study. J Clin Med. 2019;8(9):1280.

    Article  PubMed  PubMed Central  Google Scholar 

  27. Hansen NA, Wille S, Kern M. Effect of reduced airborne-particle abrasion pressure on the retention of zirconia copings resin bonded to titanium abutments. J Prosthet Dent. 2020;124(1):60–7.

    Article  PubMed  Google Scholar 

  28. Kern M. Controlled airborne-particle abrasion of zirconia ceramic restorations. J Prosthet Dent. 2010;103(2):127–8.

    Article  PubMed  Google Scholar 

  29. Happe A, Roling N, Schafer A, Rothamel D. Effects of different polishing protocols on the surface roughness of Y-TZP surfaces used for custom-made implant abutments: a controlled morphologic SEM and profilometric pilot study. J Prosthet Dent. 2015;113(5):440–7.

    Article  PubMed  Google Scholar 

  30. Schindelin J, Arganda-Carreras I, Frise E, Kaynig V, Longair M, Pietzsch T, et al. Fiji: an open-source platform for biological-image analysis. Nat Methods. 2012;9(7):676–82.

    Article  PubMed  Google Scholar 

  31. C. Sommer CS, U. Köthe and F. A. Hamprecht. Ilastik: interactive learning and segmentation toolkit. IEEE International Symposium on Biomedical Imaging: From Nano to Macro. 2011.

  32. Kern M. On the scientific evidence that the sterilization of customized implant abutments is required. Eur J Oral Implantol. 2015;8(3):219.

    PubMed  Google Scholar 

  33. Aronsson BO. On Argon-plasma cleaning—some comments from regulatory and scientific perspectives. Eur J Oral Implantol. 2015;8(3):211–2.

    PubMed  Google Scholar 

  34. Esposito M. On cleaning and sterilisation of customised abutments and disappearing implant failures. Eur J Oral Implantol. 2013;6(3):211.

    PubMed  Google Scholar 

  35. Kieschnick A. Zahntechnische Aufbereitungs- und Reinigungsverfahren von Implantat-Aufbauten. EADT e.V.; 2017.

  36. Farronato D, Fumagalli D, Asa’ad F, Rasperini G. Decontamination of customized laser-microtextured titanium abutments: a comparative in vitro study of different cleaning procedures. Int J Periodontics Restorative Dent. 2018;38(5):e87–95.

    Article  PubMed  Google Scholar 

  37. Gehrke P, Riebe O, Fischer C, Weinhold O, Dhom G, Sader R, et al. Microbiological cleaning and disinfection efficacy of a three-stage ultrasonic processing protocol for CAD-CAM implant abutments. J Adv Prosthodont. 2022;14(5):273–84.

    Article  PubMed  PubMed Central  Google Scholar 

  38. Garcia B, Camacho F, Peñarrocha D, Tallarico M, Perez S, Canullo L. Influence of plasma cleaning procedure on the interaction between soft tissue and abutments: a randomized controlled histologic study. Clin Oral Implants Res. 2016;28:1269.

    Article  PubMed  Google Scholar 

  39. Piest C, Wille S, Strunskus T, Polonskyi O, Kern M. Efficacy of plasma treatment for decontaminating zirconia. J Adhes Dent. 2018;20(4):289–97.

    PubMed  Google Scholar 

  40. Wiedenmann F, Liebermann A, Spintzyk S, Eichberger M, Stawarczyk B. Influence of different cleaning procedures on tensile bond strength between zirconia abutment and titanium base. Int J Oral Maxillofac Implants. 2019;34(6):1318–27.

    Article  PubMed  Google Scholar 

  41. Aronsson BO, Lausmaa J, Kasemo B. Glow discharge plasma treatment for surface cleaning and modification of metallic biomaterials. J Biomed Mater Res. 1997;35(1):49–73.

    Article  PubMed  Google Scholar 

  42. Gehrke P, Smeets R, Gosau M, Friedrich RE, Madani E, Duddeck D, et al. The influence of an ultrasonic cleaning protocol for CAD/CAM abutment surfaces on cell viability and inflammatory response in vitro. In Vivo. 2019;33(3):689–98.

    Article  PubMed  PubMed Central  Google Scholar 

  43. Schwendicke F, Singh T, Lee JH, Gaudin R, Chaurasia A, Wiegand T, et al. Artificial intelligence in dental research: checklist for authors, reviewers, readers. J Dent. 2021;107: 103610.

    Article  PubMed  Google Scholar 

  44. Dohan Ehrenfest DM, Coelho PG, Kang BS, Sul YT, Albrektsson T. Classification of osseointegrated implant surfaces: materials, chemistry and topography. Trends Biotechnol. 2010;28(4):198–206.

    Article  PubMed  Google Scholar 

  45. Park JH, Olivares-Navarrete R, Baier RE, Meyer AE, Tannenbaum R, Boyan BD, et al. Effect of cleaning and sterilization on titanium implant surface properties and cellular response. Acta Biomater. 2012;8(5):1966–75.

    Article  PubMed  Google Scholar 

  46. Nothdurft FP, Fontana D, Ruppenthal S, May A, Aktas C, Mehraein Y, et al. Differential behavior of fibroblasts and epithelial cells on structured implant abutment materials: a comparison of materials and surface topographies. Clin Implant Dent Relat Res. 2015;17(6):1237–49.

    Article  PubMed  Google Scholar 

  47. Al Rezk F, Trimpou G, Lauer HC, Weigl P, Krockow N. Response of soft tissue to different abutment materials with different surface topographies: a review of the literature. Gen Dent. 2018;66(1):18–25.

    PubMed  Google Scholar 

  48. Mishra PK, Wu W, Rozo C, Hallab NJ, Benevenia J, Gause WC. Micrometer-sized titanium particles can induce potent Th2-type responses through TLR4-independent pathways. J Immunol. 2011;187(12):6491–8.

    Article  PubMed  Google Scholar 

  49. Mehl C, Gassling V, Schultz-Langerhans S, Açil Y, Bähr T, Wiltfang J, et al. Influence of four different abutment materials and the adhesive joint of two-piece abutments on cervical implant bone and soft tissue. Int J Oral Maxillofac Implants. 2016;31(6):1264–72.

    Article  PubMed  Google Scholar 

  50. Nakajima K, Odatsu T, Shinohara A, Baba K, Shibata Y, Sawase T. Effects of cleaning methods for custom abutment surfaces on gene expression of human gingival fibroblasts. J Oral Sci. 2017;59(4):533–9.

    Article  PubMed  Google Scholar 

  51. Albrektsson T, Dahlin C, Jemt T, Sennerby L, Turri A, Wennerberg A. Is marginal bone loss around oral implants the result of a provoked foreign body reaction? Clin Implant Dent Relat Res. 2014;16(2):155–65.

    Article  PubMed  Google Scholar 

  52. Trindade R, Albrektsson T, Tengvall P, Wennerberg A. Foreign body reaction to biomaterials: on mechanisms for buildup and breakdown of osseointegration. Clin Implant Dent Relat Res. 2016;18(1):192–203.

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

We thank Carsten Fischer and Bredent GmbH & Co. KG for providing the Finevo cleaning system. We would also like to thank Dr Konrad Neumann (Institute of Biometry and Clinical Epidemiology, Charité-Universitätsmedizin Berlin) for his expert assistance with the statistical analysis.

Funding

Open Access funding enabled and organized by Projekt DEAL. The Oral Reconstruction Foundation (Basel, Switzerland, Grant No. ORF11804) and the CleanImplant Foundation (Berlin, Germany, Grant No. CIF2101) financially supported this study.

Author information

Authors and Affiliations

Authors

Contributions

DD and FB designed the study; PH, DD, and AK devised the methodology; AK and PH prepared the samples; PH performed the spectroscopic analyses and analysed the data with the help of DD and FS. PH wrote the manuscript and designed the figures; DD, AK, FS, and FB helped interpret the results and worked on the manuscript; FB supervised the project. All authors provided critical feedback and contributed to the research, analysis, and manuscript design.

Corresponding author

Correspondence to Paul Hofmann.

Ethics declarations

Ethics approval and consent to participate

Not applicable.

Consent for publication

Not applicable.

Competing interests

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Hofmann, P., Kunz, A., Schmidt, F. et al. Influence of exposure of customized dental implant abutments to different cleaning procedures: an in vitro study using AI-assisted SEM/EDS analysis. Int J Implant Dent 9, 33 (2023). https://doi.org/10.1186/s40729-023-00498-8

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s40729-023-00498-8

Keywords