In vivo measurement of three-dimensional load exerted on dental implants: a literature review

Abstract

Background

For biomechanical consideration of dental implants, an understanding of the three-dimensional (3D) load exerted on the implant is essential, but little information is available on the in vivo load, including the measuring devices.

Purpose

This review aimed to evaluate studies that used specific load-measuring devices that could be mounted on an implant to measure the functional load in vivo.

Materials and methods

An electronic search utilizing the internet research databases PubMed, Google Scholar, and Scopus was performed. The articles were chosen by two authors based on the inclusion and exclusion criteria.

Results

In all, 132 studies were selected from the database search, and 16 were selected from a manual search. Twenty-three studies were finally included in this review after a complete full-text evaluation. Eleven studies were related to the force measurements using the strain gauges, and 12 were related to the piezoelectric force transducer. The principles of the two types of devices were completely different, but the devices produced comparable outcomes. The dynamics of the load magnitude and direction on the implant during function were clarified, although the number of participants in each study was small.

Conclusions

The load exerted on the implant during function was precisely measured in vivo using specific measuring devices, such as strain gauges or piezoelectric force transducers. The in vivo load data enable us to determine the actual biomechanical status in more detail, which might be useful for optimization of the implant prosthetic design and development of related materials. Due to the limited data and difficulty of in vivo measurements, the development of a new, simpler force measurement device and method might be necessary.

Background

The excessive load exerted on an implant has been considered one of the primary factors causing peri-implant bone resorption, and the harmfulness of so-called “overload” has been discussed since the 1990s [1,2,3]. Thereafter, various studies reported that the load required for peri-implant resorption was beyond the level of normal occlusal force [4,5,6]. Therefore, a very low possibility of overload-related peri-implant bone resorption is generally presumed.

On the other hand, an appropriate magnitude of load is generally essential to maintain or increase bone density [7, 8]. Indeed, an important role of the functional load on the implant in the long-term health of the peri-implant bone has been suggested [4, 9]. However, bone remodelling activities might depend on the individual, location, and other anatomic and physiological parameters [10, 11]. In particular, some clinical cone beam computed tomography (CBCT) follow-up studies suggested that load-related peri-implant bone remodelling might be related to the original bone being thin [12, 13]. Therefore, the nature of the effects of load on the peri-implant bone still needs to be well understood to obtain the long-term health of peri-implant bone.

Namely, it is essential to know the actual load exerted on the implant during relevant activities to better elucidate the influence of load on peri-implant bone remodelling. Several devices have been used both clinically and experimentally to measure the actual load on the implant in vivo [14]. In most of those studies, the static occlusal force was measured using a sheet-type sensor covering full dentition (i.e., Dental Prescale (GC, Tokyo, Japan)) [15] or a biting-type sensor covering a limited area in the dentition (i.e., GM10 occlusal force-meter (Nagano Keiki, Tokyo, Japan)) [16] by engaging those devices between the upper and lower dentition. However, those methods can only measure the load magnitude at one time fragment during clenching when biting. Of note, it is more important to know the 3-dimensional (3D) real-time load exerted on the implant during functions, such as chewing, under natural occlusal conditions.

A specific high-precision small load-measuring device must be installed into the structure of implant-supported prosthetic devices to measure the load on the implant during various oral functions. Furthermore, the measuring device itself must be in a form that does not impair oral function so that the functional load on the implant can be measured in real time. Normally, engineering expertise is indispensable for developing such high-precision devices. Therefore, load measurements on implants in vivo have been very challenging, and few studies have been carried out. This study reviewed the articles that performed in vivo measurements of the load imposed on implants using specific devices and summarized the measured load data and clinical findings; these data will be useful in clinical practice and related biomechanical and mechanobiological research.

Methods

Literature search strategy

An electronic search was performed utilizing the internet research databases “PubMed”, “Google Scholar”, and “Scopus”. The search terms were constructed from the following terms: ((Dental implant) OR (Abutment)) AND ((Bite force) OR (Occlusal force) OR (Loading) OR (Mastication)) AND (In vivo) AND (Measurement)). A manual search was performed in addition to these database searches by checking the bibliography of all identified articles for potentially relevant additional studies. The search and selection process were concluded on 14 March 2022.

Inclusion and exclusion criteria

The inclusion criteria established that the studies must record an in vivo load measurement and involve the usage of a force measuring device designed to measure force on each implant prosthesis. Additionally, the full text of the studies was required to be available in English. Studies were excluded from this review if they were in vitro or animal studies and used the methodology involving the external biting-type load-measuring devices, or the full text was not available.

Study selection

Figure 1 demonstrates the literature search strategy used in this study. Two authors (IA and NY) evaluated the literature search results. First, the collected titles and abstracts were selected according to the aim and predetermined criteria. Second, two authors confirmed the concurrence of the results, and the full text of these articles was read to further examine the details of the results reported. Subsequently, the discrepancies in the results from the two authors were discussed with other authors (MI and TS). Finally, the studies that satisfied the inclusion criteria were included.

Fig. 1
figure 1

Chart of the article selection process

Results

Search results

A total of 132 studies were matched with the search terms, and 16 studies were selected from the manual hand search results by two authors (IA and NY). The abstracts and titles of all articles were screened. Unrelated titles and articles with nonclinical force measurement experimental methodology were excluded.

Finally, 23 studies were selected (by IA and NY) for inclusion in the present review. Any discrepancies in the results from the two authors were discussed with other authors (MI and TS). It was found that either strain gauges or a piezoelectric force transducer was utilized as a tool to measure the load on the implant in all selected studies. Eleven studies were related to force measurements using strain gauges, and 12 were related to measurements using the piezoelectric force transducer. Studies not included (n = 125) in the review after reading the full texts and reasons for exclusion are listed in Additional file 1: Table S1.

Studies of in vivo force measurement using a strain gauge-based device

The details of the load magnitude and moments demonstrated in each study are summarized in Table 1. The strain gauge-based load-measuring system was used to measure the axial load and bending moment exerted on the implant and/or surrounding structures during clenching or mastication in humans.

Table 1 Studies of in vivo force measurements using a strain gauge-based device

Basically, multiple strain gauge units need to be adhered onto an implant abutment surface to acquire an accurate load pattern. Hobkirk et al. used 2 units of strain gauges on each implant and reported that the load on the implant depended on the superstructure materials [17].

In several studies, 4 units of strain gauges were placed perpendicular to each other on their long axes and positioned 90 degrees to each other [18, 25, 26]. However, some studies have used 3 strain gauges placed 120 degrees from each other [19, 22, 23]. Richter et al. only used 1 unit [20], while their subsequent study and other researchers used 2 units of strain gauges on each implant [21]. Duyck et al. [22] applied the 3 strain gauge-based load-measuring system to multiple implants supporting fixed partial dentures to measure the load distribution and bending moments. The authors clarified the influence of the superstructure materials on the force distribution. A software program for analysing the load that makes the device user-friendly and simple was also updated [23]. They revealed that the distal extensions exerted greater compressive force on the closest implant, and higher forces were observed with a decreasing number of supporting implants.

Another experiment [24] in which 3 strain gauges were attached on the abutment in a single implant-supported prosthesis revealed the load magnitude during clenching and 3D chewing motions. The authors clarified that the different occlusal materials did not lead to different forces generated on the implants and the load on the implants varied across the patients, and surprisingly over 2000 N was detected in a certain patient. Another group measured the strain of the superstructure using strain gauges placed under the pontic in the case of a bride type of implant-supported 3-unit fixed partial denture, and an underestimated larger force magnitude was observed with this approach compared to that obtained with gauges attached directly to the abutment [25]. The measured load mainly depended on the location of the force application site and greater bending occurred when the force was applied at the centre of pontic compared to applying it near the abutment. Hence, a reliable measurement with the pontic method would require that a load is always applied above the strain gauge, which might result in more or less misleading force values when the loading position changes, such as when chewing foods. The method in which the strain gauges are attached to the abutment might be a more suitable application for recording measurements during natural mastication. The same group [26] analysed the effect of the occlusal surface condition of fixed prostheses, namely, the cusp inclination, on the masticatory force exerted on the supporting implants. The load exerted on the implant was also measured and compared among different materials of superstructures using similar devices with strain gauges [27]. The authors described that no significant differences in normalized bending moments were observed among the different materials; however, the load on the implants differed significantly across the patients.

Studies of in vivo force measurement using a piezoelectric transducer-based device

The details of the load magnitude and directions demonstrated in each study are summarized in Table 2. For the application of the small type of piezoelectric transducer, which was involved in the stainless-steel housing, into the dental implant, to the best of our knowledge, only two research groups have been used for the measurement of the load on the implant in humans. Mericske-Stern et al. [28] from Bern University developed an in vivo measurement system for the load on implants with piezoelectric transducers (Kistler Instruments, Winterthur, Switzerland). The device was mounted directly on the tissue-level implant (Standard RN, Straumann, Basel, Switzerland). This transducer enabled us to complete a series of studies based on in vivo measurement of the load exerted on the implants supporting an overdenture. The measured forces on the implant supporting a complete overdenture during clenching or chewing were greater than those measured during light tapping or grinding [28]. The authors also investigated the effect of distal bar extensions on the load on the implants and revealed that the distal bar extension reduced the load-sharing effect of bars [29]. They also clarified that the use of retentive ball anchorage reduced the load distributed to the implants compared to other attachments. In addition, chewing function resulted in a more pronounced transverse force component, particularly in the anterior direction, which exceeded the vertical force [30, 31]. A point of consideration was raised in that the preload of mounting the transducer on the implant could affect the reliability of the measurement. With a decrease in preload from its optimal value, the magnitude measured may be slightly reduced [32]. The authors concluded that a preload of ≥ 300N should be achieved in in vivo studies.

Table 2 Studies of in vivo force measurements using a piezoelectric transducer-based device

Another group used a piezoelectric force transducer (Kistler Instrument, Winterthur, Switzerland) with the in vivo load measurement system from Tohoku University in Japan. The authors [33, 34] investigated the load distribution between 2 implants supporting a fixed partial denture and revealed the effects of splinting the superstructure on the load distribution to both implants. Shigemitsu et al. [35, 36] used the same transducers and revealed that the load on each implant was smaller in a 4 implant-supported mandibular complete overdenture than that in 2 implant-supported type. Kobari et al. [37] applied this system to a 3-unit implant-supported fixed partial denture and investigated the effects of different numbers and configurations of supporting implants on the load distribution. The authors clarified the dynamic changes in the 3-dimensional load magnitude and direction during relevant functions in real time and the load on the implant of the cantilevered bridge type was significantly larger than the conventional bridge type. The authors also clarified that the load on the implant at gum chewing was basically larger than that at maximum voluntary clenching (MVC). Other studies [38, 39] reported that the load magnitude was affected by food texture, and the load magnitude during chewing was larger than that during MVC. The data were subsequently used for advanced experimental and numerical purposes, such as finite element analysis (FEA), and the effect of peri-implant bone resorption on the stress distribution in implant body was investigated [40].

Discussion

Among various studies regarding in vivo measurement of load on implants, this review paper focused on those in which the load was measured during function, such as clenching and mastication, in a natural occlusal condition by incorporating a specific sensing device into the implant components or superstructure. Due to their suitable size, high accuracy, wide measuring range, and high adaptability, both strain gauge-based devices and piezoelectric force transducers have been utilized for such measurements thus far.

For the measurement using the strain gauges, multiple strain gauge units are needed to acquire an accurate load pattern. Glantz et al. [41] used the Rosette strain gauge, a formation of 3 strain gauges in various orientations, to measure force on a dental bridge in multiple directions. Brunski et al. [42] developed an epoch-making load-measuring system consisting of a custom-made abutment with 3 strain gauges attached. This system was further used to quantify the amount of load applied to the implant by measuring the amount of distortion of the abutment under functional load in prosthesis-related experiments [43]. On the other hand, the piezoelectric force transducer involves some piezoelectric element materials that generate an electric charge when the load is applied. The charge is then amplified by a specific amplifier in the circuit and translated into an output voltage proportional to the load exerted on the piezoelectric crystal. The piezoelectric force transducer was first introduced for use as a bite force measuring device in the dental field by Graf et al. [44]. By installing the transducer including 3 piezoelectric elements in a fixed partial bridge, 3-dimensional forces applied to the bridge were measured. Although the principles of both methods are completely different, the outcomes are comparable to each other, and an ongoing debate about which method is superior remains. Considering the real-time accurate sensing and fast responsiveness, the piezoelectric force transducer might be more suitable for measuring the functional load on the implant in vivo.

Load data measured in vivo that quantitatively demonstrated the changes in the load magnitude and direction in real time enabled us to know the actual biomechanical status during oral function in more detail. In the 3-unit implant-supported fixed prosthesis, for example, a remarkably large load, greater than 600 N, was exerted on the implant adjacent to the pontic when using the cantilevered bridge-type superstructure during gum chewing, although the patient was an elderly woman [37]. Ducky et al. also described the increased load on the closest implant to the cantilever extension [23]. Unlike the static load to the implant during clenching, dynamic loads from various directions were also found to be exerted on the implant during chewing. The dynamic changes in the magnitude and direction of the load on the implant might be due to not only the food properties and texture [38] but also to the control mechanism and the configuration of the masticatory components [45, 46]. Three factors affecting the load directional change during functions were suggested: the friction between the two contacting teeth, the contact sliding along the incline planes of the cusp, and the formation of new contacts and loss of existing contacts influenced by the periodontal ligament. However, these findings and suggestions were obtained from a simulation study [47]. The load exerted during bruxism might be an unexpectedly large load, thus causing some mechanical complications [48], and the magnitude of intraosseous stress during bruxism might destroy osseointegration [49, 50]. Accumulation of load data measured in vivo is essential, and in particular, the detailed load data during such parafunctional habits should be valuable not only for the prevention of various biomechanical and biological complications regarding the osseointegrated implant, but also for optimization of the implant prosthetic design and related material development.

On the other hand, investigating the behaviour of the load on the implant compared to that on the natural tooth with the periodontal ligament is also important. Bassit et al. [24] revealed that a difference in resilience between acrylic resin and ceramic veneering materials used as the implant superstructure, but that difference was only measurable in in vitro where the force was generated by a shock only and the implant was rigidly anchored. The authors considered that the resilience of the masticatory system was overriding the difference in the stiffness of the various materials. The resilience in the periodontal ligament of any antagonist natural tooth might also be affected. The load on the tooth is controlled by a physiological neuromuscular mechanism, such as the muscle spindle and periodontal-masseteric reflex, during mastication [51, 52]. However, mastication is normally carried out smoothly on implant prostheses, despite the lack of periodontal ligaments, which suggests some adaptive mechanisms related to tooth loss and replacement with implants, such as activation of periodontal ligament function of opposing teeth, elicitation of neuroplasticity, and activation of osseoperception [47, 53,54,55]. In vivo measured load data will greatly contribute to the elucidation of these mechanisms.

The measured force on the implant may be affected by not only the tasks but also individual factors, such as facial morphology or masticatory muscle force. In vivo load measurement will be valuable to understand the effects of individual factors on the loads. Considering the large individual differences in human bite force [56], the wide range of measured force magnitudes among the patients (63–2055N [24]) might be reasonable. However, no studies used the strain-gauge system described the output calibration of the measuring device for that range. On the other hand, in the studies using the piezoelectric transducers, the load-measuring device was calibrated for the specified measurement range [32, 57]; thus, the measured load data might be more reliable. Clearly, the implant load changed not only vertically but also three-dimensionally during mastication and clenching [37]. In particular, for the cantilever configuration, the load increased in both the vertical and lateral directions. This excessive load might cause some mechanical complications, such as screw loosening and fracture. Therefore, we must consider the implant configuration from a biomechanical perspective in patients with multiple implants.

To date, the durability of the implant components against occlusal load and force-related peri-implant bone changes have been verified in numerous experimental studies [58]. Most of those experiments used fairly simplified loading conditions, i.e., static loading or cyclic loading tests to the implant due to the limited data of the load measured in vivo, which can be used as a reference. This might have been a general limitation of those experimental studies. To obtain more realistic results in in vitro or in silico simulation studies, reproducing the actual in vivo loading conditions as shown in this review as much as possible should be effective.

Some studies utilized in vivo measured load data for in silico simulation studies. Bing et al. [40] reported an increased risk of implant fracture associated with the amount of bone resorption due to peri-implantitis using an FEA of in vivo measured loads. The in vivo load data have also already been evaluated using FEA to investigate the distribution of peri-implant bone in a stimulation [34,35,36]. Indeed, the reliability of the simulation result was substantially improved using the in vivo measured value. Furthermore, those FEAs were utilized to construct algorithms analysing peri-implant bone remodelling [59, 60]. By comparing the results from the FEA with actual clinical outcomes of peri-implant bone changes measured in computed tomography (CT) images over time, the patient-specific bone remodelling algorithm was constructed [60]. The patient-specific bone remodelling algorithm can be extremely important data for the basis of personalized medicine in implant treatment.

A common limitation of the in vivo measurement experiments in any study introduced in this review was the small number of human participants. This is thought to be due to the cost of preparing each load-measuring device. Additionally, mounting the device to the implant components or superstructure for measuring the load under the natural occlusal condition, unlike the method of just biting the device, such as the Dental prescale, takes a large amount of time and requires great effort. Elaborate parts and precise work are essential for accurate measurement devices, which are challenging to prepare in large numbers. Difficulty securing participants was also expected. New measurement methods that replace the previous ones using strain gauges or piezoelectric transducers introduced in this review will thus be needed. We were unable to statistically analyse the loading status on implant because of the high degree of heterogeneity in load-measuring methods and the tasks performed during the measurement. Furthermore, the load-measuring devices introduced in this review are incapable of measuring the stress distribution not only in the implant components but also in the peri-implant bone as an important effect of the loading on living organisms. However, as mentioned above, the actual in vivo load data should be useful for applications in computational aided engineering methods such as FEA for investigating those stress distributions.

Conclusions

An understanding of the functional load exerted on dental implants is essential to obtain long-term favourable treatment outcomes. This review demonstrated that for the in vivo measurement of the load on implants, attaching strain gauges to the implant components and using piezoelectric force transducers have been the mainstream methods. Although both approaches have demonstrated the actual measured load on the implant during function, 3D quantitative load measurement could be carried out by using the method with a piezoelectric transducer. The development of new measurement methods is also expected to accumulate much more patient data.

Availability of data and materials

The data sets generated and analysed during the current study are available from the corresponding author upon reasonable request.

Abbreviations

CBCT:

Cone beam computed tomography

3D:

Three-dimensional

MVC:

Maximum voluntary clenching

FEA:

Finite element analysis

References

  1. Isidor F. Loss of osseointegration caused by occlusal load of oral implants. A clinical and radiographic study in monkeys. Clin Oral Implants Res. 1996;7:143–52.

    Article  PubMed  Google Scholar 

  2. Hoshaw SJ, Brunski JB, Cochran GVB. Mechanical loading of Brånemark implants affects interfacial bone modeling and remodeling. Int J Oral Maxillofac Implants. 1994;9:345–60.

    Google Scholar 

  3. Isidor F. Histological evaluation of peri-implant bone at implants subjected to occlusal overload or plaque accumulation. Clin Oral Implants Res. 1997;8:1–9.

    Article  PubMed  Google Scholar 

  4. Naert I, Duyck J, Vandamme K. Occlusal overload and bone/implant loss. Clin Oral Implants Res. 2012;23:95–107.

    Article  PubMed  Google Scholar 

  5. Kan JPM, Judge RB, Palamara JEA. In vitro bone strain analysis of implant following occlusal overload. Clin Oral Implants Res. 2012;25:73–82.

    Article  Google Scholar 

  6. Chang M, Chronopoulos V, Mattheos N. Impact of excessive occlusal load on successfully-osseointegrated dental implants: a literature review. J Investig Clin Dent. 2013;4:142–50.

    Article  PubMed  Google Scholar 

  7. Frost HM. Bone’s mechanostat: a 2003 update. Anat Rec. 2003;275:1081–101.

    Article  Google Scholar 

  8. Summer D. Long-term implant fixation and stress-shielding in total hip replacement. J Biomech. 2015;48:797–800.

    Article  Google Scholar 

  9. Kuroshima S, Nakano T, Ishimoto T, Sasaki M, Inoue M, Yasutake M, Sawase T. Optimally oriented grooves on dental implants improve bone quality around implants under repetitive mechanical loading. Acta Biomater. 2017;15:433–44.

    Article  Google Scholar 

  10. Hadjidakis DJ, Androulakis II. Bone remodeling. Ann N Y Acad Sci. 2006;1092:385–96.

    Article  PubMed  Google Scholar 

  11. Bakalova LP, Andreasen CM, Thomsen JS, Brüel A, Hauge EM, Kiil BJ, Delaisse JM, Andersen TL, Kersh ME. Intracortical bone mechanics are related to pore morphology and remodeling in human bone. J Bone Miner Res. 2018;33:2177–85.

    Article  PubMed  Google Scholar 

  12. Veltri M, Ekestubbe A, Abrahamsson I, Wennström JL. Three-dimensional buccal bone anatomy and aesthetic outcome of single dental implants replacing maxillary incisors. Clin Oral Implants Res. 2015;27:956–63.

    Article  PubMed  Google Scholar 

  13. Vera C, de Kok IJ, Chen W, Reside G, Tyndall D, Cooper LF. Evaluation of post-implant buccal bone resorption using cone beam computed tomography: a clinical pilot study. Int J Oral Maxillofac Implants. 2012;27:1249–57.

    PubMed  Google Scholar 

  14. Gu Y, Bai Y, Xie X. Bite force transducers and measurement devices. Front Bioeng Biotechnol. 2021;9: 665081.

    Article  PubMed  PubMed Central  Google Scholar 

  15. Shiga H, Komino M, Uesugi H, Sano M, Yokoyama M, Nakajima K, Ishikawa A. Comparison of two dental prescale systems used for the measurement of occlusal force. Odontology. 2020;108:676–80.

    Article  PubMed  Google Scholar 

  16. Serra CM, Manns AE. Bite force measurements with hard and soft bite surfaces. J Oral Rehabil. 2013;40:563–8.

    Article  PubMed  Google Scholar 

  17. Hobkirk JA, Psarros KJ. The influence of occlusal surface material on peak masticatory forces using osseointegrated implant-supported prostheses. Int J Oral Maxillofac Implants. 1992;7:345–52.

    PubMed  Google Scholar 

  18. Glantz PO, Rangert B, Svensson A, Stafford GD, Arnvidarson B, Randow K, et al. On clinical loading of osseointegrated implants. A methodological and clinical study. Clin Oral Implants Res. 1993;4:99–105.

    Article  PubMed  Google Scholar 

  19. Gunne J, Rangert B, Glantz PO, Svensson A. Functional loads on freestanding and connected implants in three-unit mandibular prostheses opposing complete dentures: an in vivo study. Int J Oral Maxillofac Implants. 1997;12:335–41.

    PubMed  Google Scholar 

  20. Richter EJ. In vivo vertical forces on implants. Int J Oral Maxillofac Implants. 1995;10:99–108.

    PubMed  Google Scholar 

  21. Richter EJ. In vivo horizontal bending moments on implants. Int J Oral Maxillofac Implants. 1998;13(2):232–44.

    PubMed  Google Scholar 

  22. Duyck J, Oostenvyck H, Sloten J, Cooman M, Puers R, Naert I. Influence of prosthesis material on the loading of implants that support a fixed partial prosthesis: in vivo study. Clin Implant Dent Relat Res. 2000;2:100–9.

    Article  PubMed  Google Scholar 

  23. Duyck J, Van Oosterwyck H, Vander Sloten J, De Cooman M, Puers R, Naert I. Magnitude and distribution of occlusal forces on oral implants supporting fixed prostheses: an in vivo study. Clin Oral Implants Res. 2000;11:465–75.

    Article  PubMed  Google Scholar 

  24. Bassit R, Lindström H, Rangert B. In vivo registration of force development with ceramic and acrylic resin occlusal materials on implant-supported prostheses. Int J Oral Maxillofac Implants. 2002;17:17–23.

    PubMed  Google Scholar 

  25. Morneburg TR, Pröschel PA. Measurement of masticatory forces and implant loads: a methodologic clinical study. Int J Prosthodont. 2002;15:20–7.

    PubMed  Google Scholar 

  26. Morneburg T, Proeschel P. In vivo forces on implants influenced by occlusal scheme and food consistency. Int J Prosthodont. 2002;16:481–6.

    Google Scholar 

  27. Kim HK, Heo SJ, Koak JY, Kim SK. In vivo comparison of force development with various materials of implant-supported prostheses. J Oral Rehabil. 2009;36:616–25.

    Article  PubMed  Google Scholar 

  28. Mericske-Stern R, Geering AH, Burgin WB, Graf H. Three-dimensional force measurements on mandibular implants supporting overdentures. Int J Oral Maxillofac Implants. 1992;7:185–94.

    PubMed  Google Scholar 

  29. Mericske-stern R. Force distribution on implants supporting overdentures: the effect of distal bar extensions. A 3-D in viva study. Clin Oral Implants Res. 1997;8:142–51.

    Article  PubMed  Google Scholar 

  30. Mericske-Stern R. Three-dimensional force measurements with mandibular overdentures connected to implants by ball-shaped retentive anchors. A clinical study. Int J Oral Maxillofac Implants. 1998;13:36–43.

    PubMed  Google Scholar 

  31. Mericske-Stern R, Venetz E, Fahrländer F, Bürgin W. In vivo force measurements on maxillary implants supporting a fixed prosthesis or an overdenture: a pilot study. J Prosthet Dent. 2000;84:535–47.

    Article  PubMed  Google Scholar 

  32. Mericske-stern R, Assal P, Buergin W. Simultaneous force measurements in 3 dimensions on oral endosseous implants in vitro and in vivo. A methodological study. Clin Oral Implants Res. 1996;7:378–86.

    Article  PubMed  Google Scholar 

  33. Yoda N, Gunji Y, Ogawa T, Kawata T, Sasaki K. In vivo load measurement for evaluating the splinting effects of implant-supported superstructures: a pilot study. Int J Prosthodont. 2013;26:143–6.

    Article  PubMed  Google Scholar 

  34. Shigemitsu R, Ogawa T, Matsumoto T, Yoda N, Gunji Y, Yamakawa Y, et al. Stress distribution in the peri-implant bone with splinted and non-splinted implants by in vivo loading data-based finite element analysis. Odontology. 2012;101:222–6.

    Article  PubMed  Google Scholar 

  35. Shigemitsu R, Yoda N, Ogawa T, Kawata T, Gunji Y, Yamakawa Y, et al. Biological-data-based finite-element stress analysis of mandibular bone with implant-supported overdenture. Comput Biol Med. 2014;54:44–52.

    Article  PubMed  Google Scholar 

  36. Sato E, Shigemitsu R, Mito T, Yoda N, Rasmussen J, Sasaki K. The effects of bone remodeling on biomechanical behavior in a patient with an implant-supported overdenture. Comput Biol Med. 2021;129: 104173.

    Article  PubMed  Google Scholar 

  37. Kobari H, Yoda N, Chen J, Kawata T, Sasaki K. An in vivo study on load distribution in different implant configurations for supporting fixed partial dentures. Int J Oral Maxillofac Implants. 2016;31:1049–57.

    Article  PubMed  Google Scholar 

  38. Yoda N, Ogawa T, Gunji Y, Vanegas JR, Kawata T, Sasaki K. Effects of food texture on three-dimensional loads on implants during mastication based on in vivo measurements. Implant Dent. 2016;25:515–9.

    Article  PubMed  Google Scholar 

  39. Zhang D, Han X, Zhang Z, Liu J, Jiang C, Yoda N, et al. Identification of dynamic load for prosthetic structures. Int J Numer Methods Biomed Eng. 2017;33:1–18.

    Article  Google Scholar 

  40. Bing L, Mito T, Yoda N, Sato E, Shigemitsu R, Han JM, et al. Effect of peri-implant bone resorption on mechanical stress in the implant body: in vivo measured load-based finite element analysis. J Oral Rehabil. 2020;47:1566–73.

    Article  PubMed  Google Scholar 

  41. Glantz PO, Strandman E, Svensson SA, Randow K. On functional strain in fixed mandibular reconstructions I. An in vitro study. Acta Odontol Scand. 1984;42(4):241–9.

    Article  PubMed  Google Scholar 

  42. Brunski JB, Hipp JA. In vivo forces on dental implants: hard-wiring and telemetry methods. J Biomech. 1984;17:855–60.

    Article  PubMed  Google Scholar 

  43. Glantz PO, Strandman E, Svensson SA, Randow K. On functional strain in fixed mandibular reconstructions II. An in vivo study. Acta Odontol Scandinavica. 1984;42(5):269–76.

    Article  Google Scholar 

  44. Graf H, Grassl HA. A method for measurement of occlusal forces in three dimension. Helv Odontol Acta. 1974;18:7–11.

    Google Scholar 

  45. Arima T, Takeuchi T, Honda K, Tomonaga A, Tanosoto T, Ohata N, et al. Effects of interocclusal distance on bite force and masseter EMG in healthy participants. J Oral Rehabil. 2013;40:900–8.

    Article  PubMed  Google Scholar 

  46. Kuo C, Takahashi M, Maki K. Relationships among occlusal force, condylar surface area, and facial patterns. Dent Med Res. 2013;33:169–77.

    Article  Google Scholar 

  47. deMoya AV, Schmidt ER, Eckert GJ, Katona TR. The effects of a periodontal ligament analogue on occlusal contact forces. J Oral Rehabil. 2021;49:316–26.

    Article  PubMed  Google Scholar 

  48. Svensson P. Patients with temporomandibular disorders. Adult Orthod. 2013;90:324–38.

    Article  Google Scholar 

  49. Qian J, Wennerberg A, Albrektsson T. Reasons for marginal bone loss around oral implants. Clin Implant Dent Relat Res. 2012;14:792–807.

    Article  PubMed  Google Scholar 

  50. Renvert S, Quirynen M. Risk indicators for peri-implantitis. A narrative review. Clin Oral Implants Res. 2015;26:15–44.

    Article  PubMed  Google Scholar 

  51. Kawata T, Yoda N, Kawaguchi T, Kuriyagawa T, Sasaki K. Behaviours of three-dimensional compressive and tensile forces exerted on a tooth during function. J Oral Rehabil. 2007;34:259–66.

    Article  PubMed  Google Scholar 

  52. Peck CC. Biomechanics of occlusion—implications for oral rehabilitation. J Oral Rehabil. 2015;43:205–14.

    Article  PubMed  Google Scholar 

  53. Song D, Shujaat S, Politis C, Orhan K, Jacobs R. Osseoperception following dental implant treatment: a systematic review. J Oral Rehabil. 2022;49:573–85.

    Article  PubMed  Google Scholar 

  54. Mishra SK, Chowdhary R, Chrcanovic BR, Brånemark PI. Osseoperception in dental implants: a systematic review. J Prosthodont. 2016;25:185–95.

    Article  PubMed  Google Scholar 

  55. Klineberg I, Murray G. Osseoperception: sensory function and proprioception. Adv Dent Res. 1999;13:120–9.

    Article  PubMed  Google Scholar 

  56. Levartovsky S, Peleg G, Matalon S, Tsesis I, Rosen E. Maximal bite force measured via digital bite force transducer in subjects with or without dental implants—a pilot study. Appl Sci. 2022;12:1544.

    Article  Google Scholar 

  57. Kawaguchi T, Kawata T, Kuriyagawa T, Sasaki K. In vivo 3-dimensional measurement of the force exerted on a tooth during clenching. J Biomech. 2007;40(2):244–51.

    Article  PubMed  Google Scholar 

  58. Coray R, Zeltner M, Özcan M. Fracture strength of implant abutments after fatigue testing: a systematic review and a meta-analysis. J Mech Behav Biomed Mater. 2016;62:333–46.

    Article  PubMed  Google Scholar 

  59. Zheng K, Yoda N, Chen J, Liao Z, Zhong J, Koyama S, et al. Effects of buccal thickness augmentation on bone remodeling after maxillary anterior implantation. Biomech Model Mechanobiol. 2019;19:133–45.

    Article  PubMed  Google Scholar 

  60. Liao Z, Yoda N, Chen J, Zheng K, Sasaki K, Swain MV, et al. Simulation of multi-stage nonlinear bone remodeling induced by fixed partial dentures of different configurations: a comparative clinical and numerical study. Biomech Model Mechanobiol. 2016;16:411–23.

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

Not applicable.

Funding

This work was partially supported by JSPS KAKENHI Grant Number JP19K10198.

Author information

Authors and Affiliations

Authors

Contributions

IA and NY were responsible for the study conception and design, acquisition of data, analysis and interpretation of data, drafting of the manuscript, and critical revision. MI and TS were responsible for the acquisition of data, analysis, and interpretation of data, and critical revision. TK, HE and KS were responsible for the study conception and design and critical revision of the manuscript. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Nobuhiro Yoda.

Ethics declarations

Ethics approval and consent to participate

Not applicable.

Consent for publication

Not applicable.

Competing interests

All authors declare that they have no competing interests.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Supplementary Information

Additional file 1: Table S1.

Studies not included (n=125) in the review after full text reading and reason for exclusion.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Assoratgoon, I., Yoda, N., Iwamoto, M. et al. In vivo measurement of three-dimensional load exerted on dental implants: a literature review. Int J Implant Dent 8, 52 (2022). https://doi.org/10.1186/s40729-022-00454-y

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s40729-022-00454-y

Keywords