Skip to main content

Comparison of the different voxel sizes in the estimation of peri-implant fenestration defects using cone beam computed tomography: an ex vivo study

Abstract

Background

To examine the influence of voxel sizes to detect of peri-implant fenestration defects on cone beam computed tomography (CBCT) images.

Materials and methods

This study performed with three sheep heads both maxilla and mandible and two types of dental implant type 1 zirconium implant (Zr40) (n = 6) and type 2 titanium implant (Ti22) (n = 10). A total of 14 peri-implant fenestrations (8 buccal surfaces, 6 palatal/lingual surface) were created while 18 surfaces (8 buccal, 10 palatal/lingual) were free of fenestrations. Three observers have evaluated the images of fenestration at each site. Images obtained with 0.75 mm3, 0.100 mm3, 0.150 mm3, 0.200 mm3, and 0.400 mm3 voxel sizes. For intra- and inter-observer agreements for each voxel size, Kappa coefficients were calculated.

Results

Intra- and inter-observer kappa values were the highest for 0.150 mm3, and the lowest in 0.75 mm3 and 0.400 mm3 voxel sizes for all types of implants. The highest area under the curve (AUC) values were found higher for the scan mode of 0.150 mm3, whereas lower AUC values were found for the voxel size for 0.400 mm3. Titanium implants had higher AUC values than zirconium with the statistical significance for all voxel sizes (p ≤ 0.05).

Conclusion

A voxel size of 0.150 mm3 can be used to detect peri-implant fenestration bone defects. CBCT is the most reliable diagnostic tool for peri-implant fenestration bone defects.

Introduction

Oral implants have become the most popular treatment choice for the replacement of missing teeth since introduced by Brånemark [1]. The “gold standard” material is titanium and its biomedical alloys, due to its long-term clinical survival rates for endosseous dental implants [2,3,4]. Even though titanium is the gold standard for dental implants various materials involving gold, stainless steel, and cobalt-chromium have been used in the dental implant industry. These materials are withdrawn from the oral implant industry because of their adverse side effects and had a low long-term survival rate [5, 6]. Zirconium is an alternative material because of its tooth-like color and its osseointegration potential. Besides, less plaque accumulation on zirconium than on titanium surfaces makes this material more attractive for the implant industry [7,8,9].

The long-term success of dental implants established on the health of soft and hard tissues [10]. Moreover, the most important issue is to have a sufficient amount of cortical bone around the implant because of the primary stability and osseointegration for the success of the implant treatment [11]. Additionally, inadequate cortical plate or amount of the bone revealed the risk of bone defects such as fenestration and dehiscence around the implants. Because of this situation, the success of implant treatment increases dramatically [12]. Bone defects that occur around dental implants may adversely affect gum health and cause some aesthetic problems. More importantly, implant failures may occur in the long-term. Thus, early detection of bone defects around dental implants is very important to prevent the problems [13]. Intraoral imaging modality can demonstrate the presence of bone in the mesial and distal region and also in the implant-bone inter-face. Bone loss after the initial insertion is most commonly seen in the buccal region because the bone is thinner in this area surrounding the implant [14]. Intraoral radiography techniques are two-dimensional imaging methods and have some limitations such as superimposition or distortion of anatomical structures. The visualization of buccolingual walls cannot demonstrate these methods [14, 15].

CBCT has been suggested as an alternative tool in implant dentistry for many procedures, including linear measurements of alveolar bone, graft planning, following-up after implant placement, or three dimensional (3D) evaluation of bone defects [16, 17]. It eliminates all disadvantages of two dimensional (2D) images such as superimposition, image distortion, or imaging buccolingual aspects of the bone [12, 13, 18, 19]. CBCT images allow the examination of images in all planes with submillimetric resolution and without distortion. This allows a more accurate examination of the bone defects around the implants [16, 20,21,22]. On the other hand, metallic objects used in dentistry such as amalgam or titanium implants can induce two kinds of artifacts described as beam hardening and streaking in the CBCT images. Both of these artifacts affect the visualization of areas and decrease the image quality dramatically.

Dark areas adjacent to high-density structures are called Beam hardening artifacts. This situation is explained with high-density materials that absorb the low-energy X-ray photons [23, 24]. Streak artifacts occur due to scattering radiation from metallic objects, and they are linked to the high-density objects and are seen as linear hyper densities extending along the width of the field [20, 25]. Zirconium implants create more artifacts than titanium implants in CBCT images [26]. Moreover, inaccuracies of inter-proximal peri-implant defect detection on CBCT are more definite in zirconium implants compared with titanium implants [27]. Hence, the results obtained from the studies with titanium implants cannot be applied to zirconium implants [28].

A voxel is the smallest part of a 3D image and it is isotropic in CBCT images. The selection of voxel size can be useful for the detection of peri-implant bone defects. This improves spatial resolution. In other words, increment of the spatial resolution can improve the ability of examination very small distances of the objects [20, 29]. However, there is no specific protocol described for the CBCT examination of peri-implant bone defects with different voxel size variation [30].

Hence, this study aimed to examine the effect of voxel sizes to detect of peri-implant fenestration defects on CBCT images.

Material and methods

This study performed with three sheep heads both maxilla and mandible and two types of dental implant (type 1 zirconium implant (n = 6) (Zeramex XT, Miami, FL, USA), type 2 titanium implant (n = 10) (DAND Dental implants, D.A.N.D. Metal Industries Ltd., Yavne, Israel). One periodontology specialist with experience placed all dental implants into the sheep heads (Fig. 1). After placed dental implants, the same operator creates artificial defects around dental implants. These simulating defects were created with high-speed equipment copious air/water spray to the cervical portion of the implant. Round shaped diamond burs (KG Sorensen, Zenith Dental Aps, Agerskov, Denmark) with a 3-mm diameter were used in buccal or palatal/lingual surfaces. In total 14 fenestrations (8 buccal surfaces, 6 palatal/lingual surface) were created. Eighteen surfaces (8 buccal, 10 palatal/lingual) were free of fenestrations. All fenestrations except one titanium and one zirconium implants created separately buccal or palatal/lingual surfaces; this fenestration in these implants created buccal and lingual/palatal surfaces. These bone defects were created without knowing the depth. After creating the defects around dental implants, the mucoperiosteal flap was repositioned carefully. These fenestration defects noted by the same operator to be used as a gold standard for the image evaluations.

Fig. 1
figure 1

Photo of the positioned implants and fenestration defects in the sheep’s jaw

Radiographic imaging

Sheep heads were fixed to the machine to provide standardization during CBCT scanning. All images obtained by Planmeca Promax 3D Max CBCT unit (Planmeca Oy, Helsinki, Finland) with following voxel size and parameters 0.075 mm3 (90 kV, 10 mA, 15 sn, 5 × 5.5 FOV, DAP value 730 mGy × cm2), 0.100 mm3 (90 kV, 10 mA, 12 sn, 5 × 5.5 FOV, DAP value 632 mGy × cm2), 0.150 mm3 (90 kV, 10 mA, 12 sn, 5 × 5.5 FOV, DAP value 584 mGy × cm2), 0.200 mm3 (90 kV, 10 mA, 12 sn, 5 × 5.5 FOV, DAP value 584 mGy × cm2), and 0.400 mm3 (90 kV, 10 mA, 6 sn, 5 × 5.5 FOV, DAP value 293 mGy × cm2). The images of fenestrations around titanium and zirconium implants obtained with five voxel scan modes are demonstrated in Figs. 2 and 3.

Fig. 2
figure 2

Axial, sagittal, and 3D slices from CBCT images of titanium implants with simulated peri-implant fenestration defects with five voxel sizes

Fig. 3
figure 3

Axial, sagittal and 3D slices from CBCT images of zirconium implants with simulated peri-implant fenestration defects with five voxel sizes

After image acquisition, all measurements performed on a 21.3-in. flat panel color active matrix thin film transistor (TFT) medical display (NEC MultiSync MD215MG, München, Germany) with a resolution of 2048 × 2560 at 75 Hz and 0.17-mm dot pitch operated at 11.9 bits. Three observers evaluated all CBCT scans.

Image analysis

Each scan was evaluated by 3 observers with different times of experience (ranged from 1.8 years to 4.7 years) on CBCT imaging. The observers were blinded to the clinical situation regarding the defect size and locations. All evaluations were carried out with the Planmeca CBCT unit’s own software (Romexis 4.3 Planmeca Oy, Helsinki, Finland). The observers were permitted to use both image enhancement and processing functions of the software. For CBCT, the images were anonymized while evaluating the fenestrations. All observers were trained to use the software and calibrate for the appearance of fenestrations.

For all imaging methods, a five-point scale was used to assess each fenestration visibility as (1) definitely absent; (2) probably absent; (3) unsure; (4) probably present; (5) definitely present. The observers were asked to define the fenestrations in CBCT images. Each image set was evaluated by 1 week intervals, and all evaluations were repeated 2 months after the initial examinations.

Examiner reliability and statistical analysis

Hence, in this study, AUC-ROC and Kappa coefficients were calculated to determine intra- and inter-observer agreements by different voxel sizes. Cohen’s Kappa values were clarified as κ < 0.00, no agreement; κ = 0.00–0.20, poor agreement; κ = 0.21–0.40, fair agreement; κ = 0.41–0.60, moderate agreement; κ = 0.61–0.80, good agreement; and κ = 0.81–1.00, very good agreement [31].

AUC values were clarified as AUC = 0.5: no discrimination; 0.5 < AUC < 0.7: poor discrimination; 0.7 ≤ AUC < 0.8: acceptable discrimination; 0.8 ≤ AUC < 0.9: excellent discrimination; AUC ≥ 0.9: outstanding discrimination [32].

A receiver operating characteristic curve, or ROC curve, is a graphical plot that illustrates the diagnostic ability of a binary classifier system as its discrimination threshold is varied. The ROC curve is created by plotting the true positive rate (TPR) against the false-positive rate (FPR) at various threshold settings. The true-positive rate is also known as sensitivity. The false-positive rate is also known as (1—specificity). AUC-ROC curve is a performance measurement for classification problem at various thresholds settings. ROC is a probability curve and AUC represents degree or measure of separability. It tells how much model is capable of distinguishing between classes. The higher the AUC, the better the model is at predicting 0 s as 0 s and 1 s as 1 s. By analogy, the higher the AUC, the better the model is at distinguishing between patients with disease and no disease. Statistical analyses were done for each image type, observer, and reading using the Mann-Whitney U test and the Kruskal-Wallis H test to determine the differences between the groups. Differences of a p value of less than 0.05 considered statistically significant.

Results

Intra-observer and inter-observer kappa coefficients are shown in Tables 1 and 2, respectively.

Table 1 Intraobserver kappa values calculated by scan modes and implant types for each observer
Table 2 Inter-observer kappa coefficients value and standard error according to scan modes and implant types for 1st and 2nd readings

The Kappa values of intra-observer agreements for the titanium implants according to voxel sizes were the highest for 0.150 mm3, 0.200 mm3 voxel sizes for all observers (very good agreement) while for the zirconium implants the kappa values were varied according to voxel sizes with the highest kappa value for 0.150 mm3 (good agreement) (Table 1). There are no significant differences between first and second readings for all observers (> 0.05).

Table 2 shows inter-observer kappa values for first and second readings of all observers; for zirconium implants, the lowest values achieved with 0.075 and 0.400 mm3 voxel sizes (poor to moderate). For titanium implants, the highest kappa value achieved for 0.150 mm3 voxel size (good to very good agreement) with a statistically significant difference for all inter-observer evaluations (Table 2). For the zirconium implant, there was also a statistically significant difference for 0.200 and 0.400 mm3 for all inter-observer evaluations (p ≤ 0.05).

AUC values were evaluated for each voxel size, implant type, and observer by using two readings (Table 3). While AUC values for titanium implants from ranged from 0.771 to 0.826 in 0.075 mm3 voxel sizes (acceptable to excellent discrimination), for 0.100 mm3 0.771 to 0.875 (acceptable to excellent discrimination), for 0.150 mm3 ranged from 0.785 to 0.917 (excellent to outstanding discrimination). AUC values 0.200 mm3 ranged from 0.771 to 0.875 and for 0.400 mm3 ranged from 0.726 to 0.819 (acceptable to excellent discrimination). In summary, titanium implants were found to have higher AUC values than zirconium implants with a significant difference for all voxel sizes and observers (p ≤ 0.05).

Table 3 AUC values according to scan modes and implant types for 1st and 2nd readings of the observers

Figure 4 shows the ROC curves drawn for observers 1, 2, and 3, respectively, for 1st and 2nd readings. The AUC value reflects the area under the ROC curve, with a higher value indicating a higher accuracy. The highest AUC values were obtained with the 0.150 mm3.

figure 4

Fig. 4

Table 4 shows the sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV) for all observers according to voxel sizes and implants. All voxel sizes had almost similar sensitivity and specificity values while in 0.075–0.400 mm3 voxel sizes positive predictive values were lower than 0.200, 0.150 mm3. Titanium implants were found higher sensitivity, specificity, PPV, and NPV values for all voxels than zirconium implants.

Table 4 Sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV) for the observers

Discussion

The absence of cortical plates around the cervical root or implant surfaces leads to alveolar defects such as fenestration. These defects reduce bone support for dental implants and teeth [33, 34]. The reason for the occurrence of defects may be related to the incorrect placement of the implant during surgery, excessive loading, and the inflammation caused by biofilm. Complications related to esthetic and hygiene occur due to the prevention of the defects of the implant surface from being completely overlapped [35].

Radiographs are essential methods for the detection of anatomical structures such as the alveolar bone. Buccal and lingual bone defects cannot examine with 2D imaging modalities. 3D imaging can be considered as a diagnostic tool [36]. Intraoral radiographs have shown 63–67% sensitivity in detection and identification of artificially created bone defects, CBCT has shown 80–100% sensitivity in previous studies [18, 36, 37].

Previous studies have been carried out using CBCT to evaluate dehiscence and fenestrations because it is reliable for evaluation of bone morphology with a 3D imaging [33, 38, 39].

Since voxel size affects diagnostic capacity in CBCT images, there are different results on these reports that investigate the effect of voxel size for detecting disease in the literature. Thus, in the current study, CBCT images with different voxel sizes were used to detect the fenestrations around dental implants.

Ganguly et al. used 0.16 mm3, 0.2 mm3, and 0.3 mm3 voxel size for maxillary and mandibular linear measurements in human cadaver heads then compared physical measurements and find that there was no difference between the linear measurement [40]. Librizzi et al. tested 0.2 mm3, 0.3 mm3, and 0.4 mm3 voxel sizes for detecting erosions of the mandibular condyle and found that 0.2 mm3 voxel size more useful [41]. Baltacioğlu et al. determined that there was no difference between 0.150 mm3, 0.200 mm3, and 0.400 mm3 voxel size for detecting recurrent caries [42], contrary to that Haiter et al. concluded that 0.125 mm3 voxel size provides more accurate results than 0.160 mm3, 0.250 mm3, and 0.36 mm3 voxel size for approximal caries lesion [43].

In this study depending on the software’s capabilities, all available voxel sizes were tested. It was already indicated that the smaller voxel sizes the larger radiation dose [44]. Therefore, a threshold for the voxel size to detect the fenestrations is thought to be deemed necessary. It was found in the present study, intra-observer agreements were the highest for 0.150 mm3 in both titanium and zirconium implants. However, when the voxel sizes decreases down to 0.075 mm3, the agreement drops in parallel to voxel size. This issue can be inter-pratede with the occurrence of artifacts around dental implants, as scattering or complete absorption of the beam can exist and be concluded with image degradation. This situation can prevent observation of the implant-bone inter-face and make it difficult to evaluate peri-implant bone defects [45, 46].

Kolsuz et al. asses to influence of voxel size with 0.080 mm3, 0.100 mm3, 0.125 mm3, 0.150 mm3, 0.160 mm3, and 0.200 mm3 for detection periodontal defects and indicated that there is no significant difference between voxel size up to 0.150 mm [36]. Similarly, Bagis et al. used 0.80 mm3, 0.125 mm3, and 0.160 mm3 voxel size and reported that for 0.80 mm3 and 0.125 mm3 voxel sizes has a greater degree of agreement than 0.160 mm [47].

In our study, we found that the highest kappa value at 0.150 mm3 and 0.200 mm3 voxel size lower kappa value at 0.400 mm3 for all observers and both readings when compare gold standard. These findings are similar to Kolsuz and Bagis’ results [34, 47].

A similar study compared two voxel sizes (0.12 mm and 0.2 mm) and scan modes (180° half scan and 360° full scan) using i-CAT NG unit in the estimation of titanium peri-implant fenestration and dehiscence defects. The authors found that 0.2 mm voxel size had slightly higher diagnostic values than the 0.12 mm voxel size, but there was not a significant difference in detecting bone defects. They concluded that both voxel sizes were similar to detect peri-implant fenestrations and defects [12].. Demirtürk et al. found a significant difference in their study between images acquired with higher resolution (0.2 mm and 0.25 mm voxel sizes) compared with those acquired with lower resolution (0.3 mm and 0.4 mm voxel sizes) with voxel sizes of 0.3 and 0.4 mm producing fewer artifacts. This result is in line with this current study which concluded as the moderate voxel sizes (0.150 or 0.200 mm3) showed higher agreement [48]. The present study showed a significant difference in detecting peri-implant fenestrations in different voxel resolutions both in zirconium and titanium implants. The highest agreement observed for 0.150 mm3 voxel resolution for both implant types. The smaller voxel sizes the higher radiation dose in line with higher artifact generation. The highest DAP for values achieved in smaller voxel sizes in same FOV. This may due to exposure time (s) and mAs in small voxel sizes. Moreover, images with higher voxel sizes have low-spatial resolutions, and this situation may affect the performance of diagnostic capabilities of the images [49]. The higher diagnostic capability of 0.150 mm3 voxel size compared to other voxel sizes can be explained by all this information.

The X-ray attenuation is different from a structure to another due to their atomic number and the presence of high atomic number materials. The higher the atomic number they have, the more artifact expression may be seen increasing the variability of the grey values, leading to a change of the image contrast and decrease the visualization of structures [50]. In order to test differences between dental implants, in this study, two different types of dental implant materials (Ti22 and Zr40) were used.

Demirtürk et al. also evaluated artifacts generated by zirconium, titanium, and titanium-zirconium alloy implants using different imaging modalities including CBCT in different voxel resolutions. They concluded that they found less artifacts for titanium and titanium-zirconium implants than zirconium implants [48]. Similarly, Bayrak et al. evaluated the same types of implants to detect of peri-implant dehiscences on CBCT images. They found a higher agreement on detection of peri-implant dehiscences for titanium implants than zirconium and both zirconium and titanium implants [51]. These results are in line with the current study which showed lower agreement for zirconium implant, which may interpreted as the nature of its material influence of the diagnosis of fenestration.

The limitations of this study were; first, we did not use different FOV sizes in the CBCT unit.

It is known that FOV size influences CBCT image quality because of the effect of scattered radiation [52]. In a recent study, Vasconcelos et al. assessed the performance of two metal artifact reduction (MAR) algorithms in cone beam computed tomography (CBCT) imaging, considering different materials, metal positions, and fields of view (FOVs). They stated that for ProMax3D, higher standard deviation of voxel gray values in the images acquired with small FOVs in comparison with medium FOVs [53]. Various FOV sizes can affect the quality of the images due to the scatter radiation. In this study, it was attempted to change the voxel sizes concerning FOV to assess the observer agreements as a combination of these.

The second limitation was the lack of optimization filters such as metal artifact reduction (MAR) or adaptive image noise optimizer (AINO) did not test in this study which can dramatically affect the images and reduce the artifacts and enhance image quality.

Bayrak et al. investigated the effect of a metal artifact reduction (MAR) algorithm and the adaptive image noise optimizer (AINO) optimization filter in the evaluation of dehiscences around implants with CBCT in a similar set-up. They concluded that both filters enhanced the ability of detection for the artificially created peri-implant dehiscences. They recommended that the combination of using both filters for detecting peri-implant dehiscences [51].

Vasconcelos et al. also performed a study to assess the effects of different scanning procedures both with and without MAR mode in a ProMax 3D CBCT unit (Planmeca Oy, Helsinki, Finland) to the artifact around zirconium implant. They found that selection MAR mode reduced to artifact value [54]. In this study, the MAR algorithm in this study also reduced the artifact but no statistically significant difference found for both observers in all scan modes. However, it should state that higher agreement Kappa values achieved with the gold standard in titanium implants for all each observer and scan modes.

Since FOV sizes and optimization filters such as MAR or AINO may also affect the diagnosis of peri-implant bone defects together with voxel sizes. Further studies should be carried out to examine the variables.

Conclusions

A voxel size of 0.150 mm3 was identified as the cut-off point for the overall detection of peri-implant fenestrations defects. CBCT should be considered the most reliable imaging modality for the diagnosis of periodontal defects for both zirconium and titanium implants.

Availability of data and materials

The data used and/or analyzed during the current study are available from the corresponding author on reasonable request.

Abbreviations

CBCT:

Cone beam computed tomography

Zr:

Zirconium

Ti:

Titanium

AUC:

Area under the curve

3D:

Three dimensional

2D:

Two dimensional

FOV:

Field of view

TFT:

Thin film transistor

ROC:

Receiver operating characteristic

TPR:

True-positive rate

FPR:

False-positive rate

PPV:

Positive predictive value

NPV:

Negative predictive value

MAR:

Metal artifact reduction

AINO:

Adaptive image noise optimizer

DAP:

Dose area product

References

  1. Brånemark PI, Hansson BO, Adell R, Breine U, Lindström J, Hallén O, Ohman A. Osseointegrated implants in the treatment of the edentulous jaw. Experience from a 10-year period. Scand J Plast Reconstr Surg. 1977;16:1–132.

    Google Scholar 

  2. Adell R, Eriksson B, Lekholm U, Brånemark PI, Jemt T. A long-term follow-up study of osseointegrated implants in the treatment of totally edentulous jaws. Int J Oral Maxillofac Implants. 1990;5:347–59.

    PubMed  Google Scholar 

  3. Jemt T, Chai J, Harnett J, Heath MR, Hutton JE, Johns RB, McKenna S, McNamara DC, van Steenberghe D, Taylor R, et al. A 5-year prospective multicenter follow-up report on overdentures supported by osseointegrated implants. Int J Oral Maxillofac Implants. 1996;11:291–8.

    PubMed  Google Scholar 

  4. Niinomi M. Mechanical properties of biomedical titanium alloy. Mat Sci Eng A. 1998;243:231–6.

    Google Scholar 

  5. Wataha JC. Materials for endosseous dental implants. J Oral Rehabil. 1996;23:79–90.

    PubMed  Google Scholar 

  6. Sykaras N, Lacopino AM, Marker VA, Triplett RG, Woody RD. Implant materials, designs, and surface topographies: Their effect on osseointegration. A literature review. Int J Oral Maxillofac Implants. 2000;15:675–90.

    PubMed  Google Scholar 

  7. Pieralli S, Kohal RJ, Lopez Hernandez E, Doerken S, Spies BC. Osseointegration of zirconium dental implants in animal investigations: A systematic review and meta-analysis. Dent Mater. 2017;34:171–82.

    PubMed  Google Scholar 

  8. Pieralli S, Kohal RJ, Jung RE, Vach K, Spies BC. Clinical outcomes of zirconium dental implants: a systematic review. J Dent Res. 2017;96:38–46.

    PubMed  Google Scholar 

  9. Sanz-Martin I, Sanz-Sanchez I, Carrillo de Albornoz A, Figuero E, Sanz M. Effects of modified abutment characteristics on peri-implant soft tissue health: A systematic review and metaanalysis. Clin Oral Implants Res. 2007;29:118–29.

    Google Scholar 

  10. Jones AA, Cochran DL. Consequences of implant design. Dent Clin N Am. 2006;50:339–60.

    PubMed  Google Scholar 

  11. Lekholm U, Zarb G. Patient selection and preparation. DMFR. 2013;14:40–58.

    Google Scholar 

  12. de Azevedo-Vaz SL, Vasconcelos Kde F, Neves FS, Melo SL, Campos PS, Haiter-Neto F, et al. Detection of periimplant fenestration and dehiscence with the use of two scan modes and the smallest voxel sizes of a cone-beam computed tomography device. Oral Surg Oral Med Oral Pathol Oral Radiol. 2013;115:121–7.

    Google Scholar 

  13. Haghgoo JM, Shokri A, Khodadoustan A, Khoshhal M, Rabienejad N, Farhadian M. Comparison the accuracy of the cone-beam computed tomography with digital direct intraoral radiography, in assessment of periodontal osseous lesions. Avicenna J Dent Res. 2014;6:1–6.

    Google Scholar 

  14. Schliephake H, Wichmann M, Donnerstag F, Vogt S. Imaging of periimplant bone levels of implants with buccal bone defects. Clin Oral Implants Res. 2003;14:193–200.

    PubMed  Google Scholar 

  15. Kavadella A, Karayiannis A, Nicopoulou-Karayianni K. Detectability of experimental peri-implant cancellous bone lesions using conventional and direct digital radiography. Aust Dent J. 2006;51:180–6.

    PubMed  Google Scholar 

  16. Mengel R, Kruse B, Flores-de-Jacoby L. Digital volume tomography in the diagnosis of peri-implant defects: an in vitro study on native pig mandibles. J Periodontol. 2006;77:1234–41.

    PubMed  Google Scholar 

  17. Dave M, Davies J, Wilson R, Palmer R. A comparison of cone beam computed tomography and conventional periapical radiography at detecting peri-implant bone defects. Clin Oral Implants Res. 2013;24:671–8.

    PubMed  Google Scholar 

  18. Bagis N, Kolsuz ME, Kursun S, Orhan K. Comparison of intraoral radiography and cone-beam computed tomography for the detection of periodontal defects: An in vitro study. BMC Oral Health. 2015;15:64.

    PubMed  PubMed Central  Google Scholar 

  19. Takeshita WM, Vessoni Iwaki LC, Da Silva MC, Tonin RH. Evaluation of diagnostic accuracy of conventional and digital periapical radiography, panoramic radiography, and cone-beam computed tomography in the assessment of alveolar bone loss. Contemp Clin Den. 2014;5:318–23.

    Google Scholar 

  20. Angelopoulos C, Scarfe WC, Farman AG. A comparison of maxillofacial CBCT and medical CT. Atlas Oral Maxillofac Surg Clin North Am. 2012;20:1–17.

    PubMed  Google Scholar 

  21. Corpas Ldos S, Jacobs R, Quirynen M, Huang Y, Naert I, Duyck J. Peri-implant bone tissue assessment by comparing the outcome of intra oral radiograph and cone beam computed tomography analyses to the histological standard. Clin Oral Implants Res. 2011;22:492–9.

    PubMed  Google Scholar 

  22. Sirin Y, Horasan S, Yaman D, Basegmez C, Tanyel C, Aral A, et al. Detection of crestal radiolucencies around dental implants: an in vitro,, experimental study. J Oral Maxillofac Surg. 2012;70:1540–50.

    PubMed  Google Scholar 

  23. Kamburoğlu K, Kolsuz E, Murat S, Eren H, Yüksel S, Paksoy CS. Assessment of buccal marginal alveolar peri-implant and periodontal defects using a cone beam CT system with and without the application of metal artifact reduction mode. Dentomaxillofac Radiol. 2013. https://doi.org/10.1259/dmfr.20130176.

  24. Schulze R, Heil U, Gross D, Bruellmann DD, Dranischnikow E, Schwanecke U, et al. Artefacts in CBCT: a review. Dentomaxillofac Radiol. 2011;40:265–73.

    PubMed  PubMed Central  Google Scholar 

  25. Schulze RK, Berndt D, d’Hoedt B. On cone-beam computed tomography artifacts induced by titanium implants. Clin Oral Implants Res. 2010;21:100–7.

    PubMed  Google Scholar 

  26. Sancho-Puchades M, Hämmerle CH, Benic GI. In vitro assessment of artifacts induced by titanium, titanium-zirconium and zirconium dioxide implants in cone-beam computed tomography. Clin Oral Implants Res. 2015;26:1222–8.

    PubMed  Google Scholar 

  27. Steiger-Ronay V, Krcmaric Z, Schmidlin PR, Sahrmann P, Wiedemeier DB, Benic GI. Assessment of peri-implant defects at titanium and zirconium dioxide implants by means of periapical radiographs and cone beam computed tomography: an in-vitro examination. Clin Oral Implants Res. 2018;29:1195–201.

    PubMed  Google Scholar 

  28. Liedke GS, Spin-Neto R, da Silveira HED, Schropp L, Stavropoulos A, Wenzel A. Factors affecting the possibility to detect buccal bone condition around dental implants using cone beam computed tomography. Clin Oral Implants Res. 2017;28:1082–8.

    PubMed  Google Scholar 

  29. Benavides E, Rios HF, Ganz SD, An CH, Resnik R, Reardon GT, et al. Use of cone beam computed tomography in implant dentistry: the international congress of oral implantologists consensus report. Implant Dent. 2012;21:78–86.

    PubMed  Google Scholar 

  30. Spin-Neto R, Gotfredsen E, Wenzel A. Impact of voxel size variation on CBCT-based diagnostic outcome in dentistry: a systematic review. J Digit Imaging. 2013;26:813–20.

    PubMed  Google Scholar 

  31. Jacob C. A coefficient of agreement for nominal scales. Educ Psychol Meas. 1960;20:37–46.

    Google Scholar 

  32. Fawcett T. An introduction to ROC analysis. Pattern Recogn Lett. 2006;27:861–74.

    Google Scholar 

  33. Enhos S, Uysal T, Yagci A, Veli I, Ucar FI, Ozer T. Dehiscence and fenestration in patients with different vertical growth patterns assessed with cone-beam computed tomography. Angle Orthod. 2012;82:868–74.

    PubMed  Google Scholar 

  34. Leung CC, Palomo L, Griffith R, Hans MG. Accuracy and reliability of cone-beam computed tomography for measuring alveolar bone height and detecting bony dehiscences and fenestrations. Am J Orthod Dentofac Orthop. 2010;137:109–19.

    Google Scholar 

  35. Blanco J, Alonso A, Sanz M. Long-term results and survival rate of implants treated with guided bone regeneration: a 5-year case series prospective study. Clin Oral Implants Res. 2005;16:294–301.

    PubMed  Google Scholar 

  36. Kolsuz ME, Bagis N, Orhan K, Avsever H, Demiralp KO. Comparison of the influence of FOV sizes and different voxel resolutions for the assessment of periodontal defects. Dentomaxillofac Radiol. 2015. https://doi.org/10.1259/dmfr.20150070.

  37. Misch KA, Yi ES, Sarment DP. Accuracy of cone beam computed tomography for periodontal defect measurements. J Periodontol. 2006;77:1261–6.

    PubMed  Google Scholar 

  38. Yagci A, Veli I, Uysal T, Ucar FI, Ozer T, Enhos S. Dehiscence and fenestration in skeletal Class I, II, and III malocclusions assessed with cone-beam computed tomography. Angle Orthod. 2012;82:67–74.

    PubMed  Google Scholar 

  39. Noujeim M, Prihoda T, Langlais R, Nummikoski P. Evaluation of high-resolution cone beam computed tomography in the detection of simulated interradicular bone lesions. Dentomaxillofac Radiol. 2009;38:156–62.

    PubMed  Google Scholar 

  40. Ganguly R, Ramesh A, Pagni S. The accuracy of linear measurements of maxillary and mandibular edentulous sites in cone-beam computed tomography images with different fields of view and voxel sizes under simulated clinical conditions. Imaging Sci Dent. 2016;46:93–101.

    PubMed  PubMed Central  Google Scholar 

  41. Librizzi ZT, Tadinada AS, Valiyaparambil JV, Lurie AG, Mallya SM. Cone-beam computed tomography to detect erosions of the temporomandibular joint: Effect of field of view and voxel size on diagnostic efficacy and effective dose. Am J Orthod Dentofac Orthop. 2011;140:25–30.

    Google Scholar 

  42. Baltacioglu İH, Eren H, Yavuz Y, Kamburoglu K. Diagnostic accuracy of different display types in detection of recurrent caries under restorations by using CBCT. Dentomaxillofac Radiol. 2016. https://doi.org/10.1259/dmfr.20160099.

  43. Haiter-Neto F, Wenzel A, Gotfredsen E. Diagnostic accuracy of cone beam computed tomography scans compared with intraoral image modalities for detection of caries lesions. Dentomaxillofac Radiol. 2008;37:18–22.

    PubMed  Google Scholar 

  44. Hekmatian E, Jafari-Pozve N, Khorrami L. The effect of voxel size on the measurement of mandibular thickness in cone-beam computed tomography. Dent Res J (Isfahan). 2014;11:544–8.

    Google Scholar 

  45. Benic GI, Sancho-Puchades M, Jung RE, Deyhle H, Hammerle CH. In vitro assessment of artifacts induced by titanium dental implants in cone beam computed tomography. Clin Oral Implants Res. 2013;24:378–83.

    PubMed  Google Scholar 

  46. Pauwels R, Stamatakis H, Bosmans H, Bogaerts R, Jacobs R, Horner K, Tsiklakis K, SEDENTEXCT Project Consortium. Quantification of metal artifacts on cone beam computed tomography images. Clin Oral Implants Res 2013;24(Suppl A100):94–99.

  47. Bagis N, Eren H, Kolsuz ME, Kurt MH, Avsever H, Orhan K. Comparison of the burr and chemically induced periodontal defects using different field-of-view sizes and voxel resolutions. Oral Surg Oral Med Oral Pathol Oral Radiol. 2018;125:260–7.

    PubMed  Google Scholar 

  48. Demirturk Kocasarac H, Ustaoglu G, Bayrak S, et al. Evaluation of artifacts generated by titanium, zirconium, and titanium-zirconium alloy dental implants on MRI, CT, and CBCT images: A phantom study. Oral Surg Oral Med Oral Pathol Oral Radiol. 2019;127:535–44.

    PubMed  Google Scholar 

  49. Kim DS, Rashsuren O, Kim EK. Conversion coefficients for the estimation of effective dose in cone-beam CT. Imaging Sci Dent. 2014;44:21–9.

    PubMed  PubMed Central  Google Scholar 

  50. Gerlach NL, Meijer GJ, Borstlap WA, Bronkhorst EM, Bergé SJ, Maal TJJ. Accuracy of bone surface size and cortical layer thickness measurements using cone beam computerized tomography. Clin Oral Implants Res. 2013;24:793–7.

    PubMed  Google Scholar 

  51. Bayrak S, Orhan K, Kursun Çakmak ES, Görürgöz C, Odabaşı O, Yilmaz D, Atakan C. Evaluation of a metal artifact reduction algorithm and an optimization filter in the estimation of peri-implant dehiscence defects by using cone beam computed tomography: an in-vitro study. Oral Surg Oral Med Oral Pathol Oral Radiol. 2020. https://doi.org/10.1016/j.oooo.2020.02.005.

  52. Pauwels R, Jacobs R, Bogaerts R, Bosmans H, Panmekiate S. Reduction of scatter-induced image noise in cone-beam CT: effect of field of view size and position. Oral Surg Oral Med Oral Pathol Oral Radiol. 2016;121:188–95.

    PubMed  Google Scholar 

  53. Vasconcelos KF, Codari M, Queiroz PM, et al. The performance of metal artifact reduction algorithms in cone beam computed tomography images considering the effects of materials, metal positions, and fields of view. Oral Surg Oral Med Oral Pathol Oral Radiol. 2019;127(1):71–6.

    PubMed  Google Scholar 

  54. Vasconcelos TV, Bechara BB, McMahan CA, Freitas DQ, Noujeim M. Evaluation of artifacts generated by zirconium implants in cone-beam computed tomography images. Oral Surg Oral Med Oral Pathol Oral Radiol. 2017;123:265–72.

    PubMed  Google Scholar 

Download references

Acknowledgements

Not applicable

Funding

Not applicable

Author information

Authors and Affiliations

Authors

Contributions

MHK, KO, and NB conceptualized and designed the study. NB performed the implant placement. CE acquired the data. MHK, CE, and NB observed the data. CA made the statistical analysis. MHK and KO contributed to the writing and revision of the manuscript. All authors have read and approved the manuscript.

Corresponding author

Correspondence to Mehmet Hakan Kurt.

Ethics declarations

Ethics approval and consent to participate

Not applicable

Consent for publication

Not applicable

Competing interests

Not applicable

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Kurt, M.H., Bağış, N., Evli, C. et al. Comparison of the different voxel sizes in the estimation of peri-implant fenestration defects using cone beam computed tomography: an ex vivo study. Int J Implant Dent 6, 58 (2020). https://doi.org/10.1186/s40729-020-00254-2

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s40729-020-00254-2

Keywords