From: Comparison of the accuracy of immediate implant placement using static and dynamic computer-assisted implant system in the esthetic zone of the maxilla: a prospective study
Group
Static (n = 20)
Dynamic (n = 20)
p-value
Global entry deviation (mm)
0.99 ± 0.63
1.06 ± 0.55
0.659
Global apex deviation (mm)
1.50 ± 0.75
1.18 ± 0.53
0.231
Angular deviation (°)
3.07 ± 2.18
3.23 ± 1.67
0.547