Skip to main content

Table 1 Absolute and relative (percentage) calculated and measured dose by film dosimetry for point U

From: Investigating dose homogeneity in radiotherapy of oral cancers in the presence of a dental implant system: an in vitro phantom study

U Calculated cGy (%)
Implanted Non-implanted
 M-plan 198 (99%) -
 D-plan 212 (106%) 197 (98.7%)
 W-plan 201 (100.5%) -
Isocenter   200 200
U Measured cGy (%)
Implanted* Non-implanted
  213 (98.6%) 192 (98.4%)
Isocenter   216 195
  1. *Monitor Unit was set from the default plan which no material was predefined