Study | Criteria | Abutment fracture | Abutment loosening | Prosthesis complication | Abutment fracture rate | Mechanical failure rate |
---|---|---|---|---|---|---|
Andersson et al. 2001 [15] | N/A | 5 during prep or placement 2 minor chip fractures (B) 2 after loading (1 M and 7 M) | No | crown fracture of c group at 2 Y | 7% | 8.99% |
Galluci et al. 2011b [51] | N/A | No | No | 2 minor ceramic chipping in t | No | 10.00% |
Hosseini et al. 2011 [66] | N/A | No | No | 1 ceramic chipping and 1 crown loss of retention at 1Y in metal group | No | 2.67% |
Thoma et al. 2016a [47] | USPHS | 1 in t before 1 Y | No | 1 chipping (in t) and 3 occlusal roughness (in c) at 1 Y | 5% | 25.00% |
Carrillo de Albornoz et al. 2014 [67] | N/A | 2 (in t) | No | No | 8 % | 8.00% |
Baldini et al. 2016 [68] | N/A | No | No | No | No | 0.00% |
Paolantoni et al. 2016 [69] | N/A | 2 (one piece) 1 (two-piece) | No | 3 all-ceramic (one piece) fractured (26 M) | 10.3% | 8.11% |
Wittneben et al. 2017 [58] Wittneben et al. 2020 [59] | N/A | No | No | 1 crown fracture (in t) at 1 Y 1 major ceramic chipping 1 crown replacement due to esthetics at 3 Y (in c) | No | 7.69% |
Amorfini et al. 2018 [73] | N/A | 1 in CR | 2 one in each | 1 chipping in SR 1 crown decementation | 3% | 15.63% |
Eisner et al. 2018 [49] | USPHS | No | 1 (in t) at 3Y | 2 (one in each group) minor chipping 6 (4t/2c) occlusal wear at 3 Y | 5% (same as 1Y) | 45.00% |
Thoma et al. 2018a [54] | USPHS | No | No | 2 minor chipping in CR at B 1 poor marginal adaptation in CR at B and 6M 50% contact points alpha score at B and 6M | N/A | 9.09% |
Heierle et al. 2019 [53] | USPHS | 1 in SR | No | 1 veneer fracture in SR 4 in each group occlusal wear | 3.7% | 14.71% |
Laass et al. 2019 [50] | USPHS | No | No | 12 slight occlusal roughness (5t, 7c) 2 larger occlusal roughness 2 minor chipping | 5% (same as 1Y) | 80.00% |
Kraus et al. 2019 [57] | USPHS | 6 (2 in CR and 4 in SR) | No | Slight occlusal roughness: 4 in SR and 9 in CR Minor chipping in 1 in CR | 13.6% | 45.45% |