Fig. 4 | International Journal of Implant Dentistry

Fig. 4

From: An experimental study on the effects of the cortical thickness and bone density on initial mechanical anchorage of different Straumann® implant designs

Fig. 4

a BLX implant showed higher ISQ values compared to other groups. The significant difference was found when compared to SP and BLT groups (p 0.001 and p 0.0016, respectively). Besides that, TE implants showed also statistically significant superiority compared to SP (p 0.001). The difference between BLT and SP groups was also remarkable (p 0.003). b TE implants showed slightly higher values compared to BLX; however, the difference was statistically insignificant. BLX and TE implants showed significantly higher ISQ values compared to BLT and SP groups (BLX-BLT p 0.004, BLX-SP p 0.001 and TE-BLT p 0.001, TE-SP p 0.001, respectively)