Accuracy of reference scanners | |||
---|---|---|---|
Scanner | Manufacturer | Reference | Accuracy |
Dental Laboratory Scanner | |||
 Activity 880 scanner | Smart Optics | Amin 2016 [23] Marghalani 2018 [9] | Precision: 10 μm |
 D-250(3Shape) | 3 Shape | Flugge 2016 [15] | 2 μm |
 Freedom UHD® | DOF | Mangano 2019 [24] | 5 μm |
 Iscan D103I | Imetric | Papaspiridacos 2016 | 6 μm |
 Iscan D104I | Imetric | Mangano 2016 [25] | Trueness < 5 μm, Precision < 10 μm |
 Lava Scan ST scanner | 3M | Andriessen 2014 [20] Lee 2015 | Not found |
Industrial 3D scanner | |||
 ATOS Compact Scan 5M | GOM | Arcuri 2019 [26] |  |
 ATOS Core 80 | GOM | Alikhasi 2018 [19] Nedelcu 2018 [8] | Precision 4 μm (Alikhasi 2018) [19] Precision 0.6 μm (Nedelcu 2018) [8] |
 Infinite Focus Standard | Alicona Imaging | Ender 2013 | Trueness 5.3 ± 1.1 μm, Precision of 1.6 ± 0.6 μm (Ender 2013) Manufacturer’s information: Trueness 0.5 μm, Precision 0.1 μm |
 ScanRider | V-GER | Imburgia 2017 [27] | Trueness 5–10 μm, Precision 15–30 μm |
 stereoSCAN neo | AICON 3D Systems | KimRJY 2019 [28] |  |
Coordinate measuring machine (CMM) | Â | Â | Maximum error (L = Length: mm) |
 CONTURA | Zeiss | Kim 2019 | Trueness: 1.5 + L/350 (Manufacturer’s information: Truness 0.7 μm, Precision 0.55 μm (Kim 2019) |
 Crista-Apex | Mitutoyo | Gimenez 2015 [18] Gimenez 2017 Gintaute 2018 [29] | Trueness: 1.9 + 3L/1000 μm |
 Crista Apex S | Mitutoyo | Menini 2017 | Trueness: 1.9 + 3L/1000 μm |
 DEA Mistral | Brown & Sharpe | Alikhasi 2018 [19] | Trueness: 3.5 + L/250 μm |
 Edge ScanArm HD | FARO | Sami 2020 [30] | Trueness: 25 μm, Precision: 25 μm |
 Global Silver Performance 7.10.7 | Brown & Sharpe | Chia 2017 [31] Tan 2019 [10] | Trueness: 1.9–2.0 μm |
 Leitz PMM 12106 | Zeiss | Van der Meer 2012 [13] | Trueness: 0.3 μm, Precision: 0.1 μm |
 SmartScope Flash, CNC 300 | Optical Gaging Products | Di fore 2019 | Trueness: 2.8 + 5 L/1000 μm |
 UPMC 550-CARAT | Zeiss | Ajioka 2016 [14] Fukazawa 2017 [12] | Trueness: 0.8 + L/600 μm |