Table 1 Studied CBCT devices with corresponding codes and scanning parameters

From: Impact of the blooming artefact on dental implant dimensions in 13 cone-beam computed tomography devices

CBCT devices

Code

Medium FOV standard

Small FOV standard

Small FOV high-resolution

FOV (cm)

kV

mA

Voxel size (mm)

Exposure time (s)

DAP (mGy cm2)

FOV (cm)

kV

mA

Voxel size (mm)

Exposure time (s)

DAP (mGy cm2)

FOV (cm)

kV

mA

Voxel size (mm)

Exposure time (s)

DAP (mGy cm2)

3D Accuitomo 170

A

8 × 8

90

5

0.125

30.8

2310

4 × 4

90

5

0.125

17.5

402

4 × 4

90

5

0.08

30.8

706

Veraview X800

B

8 × 8

100

8

0.125

17.9

2112

4 × 4

100

8

0.125

17.9

596

4 × 4

100

8

0.08

17.9

589

X1

C

8 × 8

90

12

0.15

8.73

773

4 × 4

90

12

0.15

8.73

231

4 × 4

90

12

0.1

11.7

309

NewTom VGi evo

D

8 × 8

110

3

0.125

4.32

329

5 × 5

110

3

0.125

4.32

148

5 × 5

110

3

0.1

4.32

148

OP 3D Pro

E

8 × 8

90

7

0.2

6

899

5 × 5

90

7

0.125

6

452

5 × 5

90

9

0.08

8.7

775

OP 3D

F

6 × 9

95

3

0.2

20

940

5 × 5

95

4

0.125

10

582

5 × 5

95

4

0.08

20

1165

CS 9300

G

8 × 8

90

4

0.18

8

455

5 × 5

90

4

0.2

20

572

5 × 5

90

4

0.09

20

572

ProMax 3D Max

H

10 × 10

96

7

0.15

15

668

6 × 6

96

7

0.15

15

657

6 × 6

96

12

0.07

15

1155

Orthophos SL 3D

I

8 × 8

85

6

0.16

14

657

5 × 5

85

10

0.16

4.3

214

5 × 5

85

6

0.08

14

421

Spectral Dental CBCT

J

9 × 9

110

8

0.2

17

462

6 × 6

110

8

0.2

17

200

6 × 6

110

8

0.07

27

527

Veraviewepocs 3D R100

K

8 × 8

90

8

0.125

9.4

1780

4 × 4

90

8

0.125

9.4

525

*

*

*

*

*

*

KaVo 3D eXam+

L

8 × 8

120

5

0.125

7.4

500

*

*

*

*

*

*

*

*

*

*

*

*

Pax-i3D Green

M

8 × 8

104

5

0.120

11

409

*

*

*

*

*

*

*

*

*

*

*

*

  1. FOV Field of view, kV kilovoltage, mA milliamperage, DAP Dose-area product indicated by the manufacturer
  2. *Protocol not available