Skip to main content

Table 1 Mean removal torque values (N cm) of the abutment screws of the implant-abutment connections in all groups

From: Removal torque pattern of a combined cone and octalobule index implant-abutment connection at different cyclic loading: an in-vitro experimental study

Group

Number of cycles

Mean removal torque value (N cm) ± SD

% of mean RTV loss compared to initial RTV

0

0

27.67 ± 1.66 (initial RTV)

1

50,000

21.77 ± 2.13

21.33%

2

100,000

20.03 ± 2.21

27.59%

3

150,000

19.45 ± 2.82

29.70%

4

200,000

17.10 ± 4.79

38.19%

5

400,000

17.37 ± 3.92

37.29%

6

600,000

19.07 ± 1.48

31.08%

7

800,000

19.93 ± 3.00

27.95%

8

1,000,000

17.78 ± 3.03

35.75%

9

1,200,000

15.99 ± 3.40

42.20%

10

1,400,000

18.28 ± 1.02

33.95%

11

1,600,000

19.04 ± 3.03

31.17%

12

1,800,000

19.16 ± 2.45

30.75%

13

2,000,000

15.84 ± 3.42

42.74%