Skip to main content

Table 2 One-way ANOVA and post hoc Tukey test results for mean ± SD of the initial RTV, postload RTV, and difference between initial and postload RTV between all groups

From: Effect of different angulations and collar lengths of conical hybrid implant abutment on screw loosening after dynamic cyclic loading

 

Initial RTV

Postload RTV

Difference between initial and postload RTV

Mean ± SD

Subgroups

 GIA

22.5 ± 0.5

23.1 ± 0.8

− 0.6 ± 0.4

 GIB

22.4 ± 0.4

22.7 ± 0.4

− 0.2 ± 0.2

 GIIA

22.4 ± 0.4

20.8 ± 0.5

1.6 ± 0.1

 GIIB

22.2 ± 0.4

19.1 ± 0.9

3.0 ± 0.5

 GIIIA

22.2 ± 0.3

18.7 ± 0.8

3.4 ± 0.5

 GIIIB

22.2 ± 0.4

17.1 ± 1.6

5.0 ± 1.2

ANOVA

F

1.5

88.0

153.9

P value

0.1

< 0.001*

< 0.001*

After value

Tukey’s test

GIA

GIB

GIIA

GIIB

GIIIA

GIB

0.8

    

GIIA

< 0.001*

< 0.001*

   

GIIB

< 0.001*

< 0.001*

< 0.001*

  

GIIIA

< 0.001*

< 0.001*

< 0.001*

0.9

 

GIIIB

< 0.001*

< 0.001*

< 0.001*

< 0.001*

< 0.001*

  1. − negative means decreased post loading RTL ratio (removal torque gain)
  2. *statistical significant differences (P <.05)