Table 2 Result of IT and ISQ

From: Prospective multicenter non-randomized controlled study on intraosseous stability and healing period for dental implants in the posterior region

Number of implants

Insertion torque value (N cm)

Implant stability quotient value

0 week

2 weeks

4 weeks

6 weeks

8 weeks

12 weeks

1

25

33.0

75.0

77.0

78.3

79.7

77.0

2

40

68.0

70.3

70.0

72.0

75.7

75.3

3

40

78.3

77.0

78.0

78.7

80.0

80.0

4

35

74.0

43.0

61.0

73.0

75.7

80.0

5

45

85.3

85.7

84.0

83.3

84.0

83.0

6

25

65.0

68.0

68.3

70.0

72.0

71.0

7

40

82.3

84.0

82.0

84.0

81.0

81.0

8

10

51.3

52.1

46.7

59.3

66.3

68.7

9

35

78.7

80.0

80.0

80.0

81.3

80.0

10

35

86.3

84.0

83.0

84.0

85.0

85.0

11

40

78.0

75.7

76.3

78.7

79.7

80.0

12

40

80.0

80.0

79.3

81.0

81.0

84.3

13

35

71.0

73.0

77.3

80.0

77.0

82.0

14

45

70.0

72.0

76.0

80.0

81.0

82.3

15

30

73.3

73.0

65.0

68.7

73.3

75.0

16

25

75.3

75.0

72.0

73.0

75.3

77.7

17

35

68.7

76.0

72.7

76.3

79.0

81.3

18

30

75.0

80.0

80.3

81.7

82.7

81.3

19

30

50.7

20

30

73.7

61.3

67.7

73.7

57.7

57.0

21

45

69.3

66.7

57.3

75.0

76.3

77.7

22

20

73.3

75.7

76.7

77.0

80.7

82.0

23

50

71.7

69.7

73.3

76.0

77.3

79.7

24

30

76.0

80.0

80.0

82.0

81.3

81.0

25

20

64.0

47.7

71.0

69.3

68.0

77.0

26

20

65.0

70.0

75.0

77.0

79.0

80.0

27

35

65.7

60.3

61.3

69.7

68.0

68.0

28

40

22.3

57.0

68.7

75.0

75.0

80.0

29

35

69.7

73.7

76.7

74.7

74.7

81.0

30

20

58.0

46.0

65.7

70.0

70.0

75.3

31

35

67.0

78.0

69.3

78.7

80.0

82.3

32

20

53.3

63.3

61.0

64.7

68.7

70.0

33

40

52.0

52.7

66.0

75.7

78.0

79.3

  1. The primary stability of no.19 implant was evaluated as good. ISQ of this implant was measured immediately after surgery. No.19 implant had mobility at 2 weeks after surgery and did not improve even at 4 weeks after surgery. Therefore, this implant was excluded from the evaluation