Skip to main content

Table 2 Mean values (± SD) of recorded maximum RF values, calculated indirect ISQ values, and direct recorded ISQ values for Ankylos (A) and Straumann (S) test implants

From: ISQ calculation evaluation of in vitro laser scanning vibrometry-captured resonance frequency

Batch #

Implant system

Implant length (mm)

Implant diameter (mm)

Mean measured resonance freq (kHz)

SD measured resonance freq (kHz)

Mean indirect ISQ

SD indirect ISQ

Mean direct ISQ

SD direct ISQ

26

A

9.5

4.5

8607.60

0.894

89.19

0.006

90

0

27

A

9.5

4.5

8000.00

0

85.02

0

85

0

28

A

9.5

4.5

8032.00

0

85.24

0

85

0

29

A

9.5

4.5

8256.00

0

86.77

0

85.8

1.0954

30

A

9.5

4.5

8256.00

0

86.77

0

87

0

40

S

12

3.3

5180.00

3.346

65.63

0.025

65

0

41

S

12

3.3

5180.00

2.828

65.62

0.0212

65

0

42

S

12

3.3

5180.00

2.828

65.62

0.0212

65

0

43

S

12

3.3

5180.00

2.828

65.62

0.0212

65

0

44

S

12

3.3

5038.40

2.190

64.54

0.0168

64

0

ts1

S

4.1

10

7257.60

26.773

79.95

0.1814

80

0

ts2

S

4.1

10

7251.20

17.5271

79.90

0.1187

80

0

ts3

S

4.1

10

7225.60

14.3108

79.93

0.0968

80

0

ts4

S

4.1

10

7206.40

14.3108

79.60

0.0968

80

0

ts5

S

4.1

10

7232.00

0

79.77

0

80

0

55

S

4.8

8

7070.80

17.922

78.68

0.1209

75

0

56

S

4.8

8

7100.00

30.4302

78.88

0.2055

79

0

57

S

4.8

8

6857.60

21.4196

77.24

0.1441

77

0

58

S

4.8

8

7375.20

41.8473

80.7448

0.2842

77

0

59

S

4.8

8

7115.20

31.5150

78.9835

0.2129

78

0