Table 2 The results of logistic regression of method used on the occurrence of perforation

From: Comparison of three different methods of internal sinus lifting for elevation heights of 7 mm: an ex vivo study

Methods

BAOFSE

BALLOON

CAS kit

Number of cases

12

12

12

Number of perforations

7

1

1

Percentage

58.4%

8.3%

8.3%

Comparison of methods regarding perforations (odds ratio)

 

Balloon\BOAFSE

Balloon\CAS kit

CAS kit\BAOFSE

Odds ratio

0.091

1

0.091

p value

0.022

1

0.022

Confidence interval

Lower

Upper

Lower

Upper

Lower

Upper

 

1.437

160.972

0.55

18.085

1.437

160.972

Reference group

BAOSFE

BAOSFE

CAS kit