Skip to main content

Table 2 Insertion torque value and removal torque value

From: Effect of implant design on primary stability using torque-time curves in artificial bone

Code

Insertion torque

Removal torque

Effective thread length (ETL)

(N · cm)

(N · cm)

(×π mm)

ST6

6.19 ± 0.716

5.95 ± 0.718

11.53

ST8

8.06 ± 1.038

9.09 ± 1.093

15.11

ST10

13.13 ± 1.763

12.37 ± 1.746

21.48

BL8

17.67 ± 1.290

16.67 ± 2.140

20.88

BL10

23.56 ± 1.628

21.99 ± 1.530

31.00

BL12

26.66 ± 3.897

24.40 ± 2.298

39.96

TE10

25.17 ± 2.374

23.76 ± 2.027

31.11

MK3

16.03 ± 0.516

10.30 ± 0.708

39.38

MK4

39.35 ± 0.494

34.31 ± 0502

54.81

 

±:SD