Skip to main content

Table 5 Three-way ANOVA (displacement in the inferior-superior direction [ z -axis])

From: A biomechanical investigation of mandibular molar implants: reproducibility and validity of a finite element analysis model

Source Sum of squares df Mean squared F value p value
Contact model
 A: Observed area 22.324 1 22.324 68.424 0.014*
 B: Dental formula 610.338 2 305.169 915.448 0.000**
 C: Loading points 92.755 2 46.377 22.619 0.007**
Fixation model
 A: Observed area 16.600 1 16.600 360.045 0.003**
 B: Dental formula 190.012 2 95.006 2641.293 0.000**
 C: Loading points 27.806 2 13.903 78.581 0.001**
  1. *p < 0.05, **p < 0.01.