Skip to main content

Table 1 Comparison of deviations when using real-time navigation and static pilot drilling template

From: Accuracy of intraoral real-time navigation versus static, CAD/CAM-manufactured pilot drilling guides in dental implant surgery: an in vitro study

n = 30/30

Navigation

Template

p-value

Significance

Angle

4.6°

0.0016

*

Apical base

0.445 mm

0.275 mm

0.003

*

Apical tip

0.43 mm

0.23 mm

0.0006

*

Distal base

0.31 mm

0.285 mm

0.7662

 

Distal tip

0.43 mm

0.48 mm

0.7888

 

Vestibular base

0.42 mm

0.185 mm

0.0079

*

Vestibular tip

0.605 mm

0.34 mm

0.0341

*

3D offset base

0.8450

0.62 mm

0.0044

*

3D offset tip

0.9950

0.8250

0.2344

 

Time

4.66 min (280 s)

3.175 min (190.5 s)

<0.0001

*